{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cortlandt Manor",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64537",
      "recalling_firm": "Polymedco, Inc",
      "address_1": "510 Furnace Dock Rd",
      "address_2": "N/A",
      "postal_code": "10567-6220",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada and Guam.",
      "recall_number": "Z-1059-2013",
      "product_description": "Test Strips packaged in the OC-Light Manual iFOBT Kit, Catalog No. FOB50.    For the rapid, qualitative detection of human hemoglobin in feces.",
      "product_quantity": "355,350 test strips",
      "reason_for_recall": "The test strips in the affected product lots of the OC Light immunological fecal occult blood test are being recalled because they may not be stable through their labeled shelf-life.  There may be a possibility of obtaining false negative results when using the affected lots.",
      "recall_initiation_date": "20130212",
      "center_classification_date": "20130404",
      "termination_date": "20160210",
      "report_date": "20130410",
      "code_info": "Kit Lot # P205877 - Kit Expiration Date 2/28/2013 - Test Strip Lot # 18004M;    Kit Lot # P210125 - Kit Expiration Date 3/31/2013 - Test Strip Lot # 19006M;    Kit Lot # P210204 - Kit Expiration Date 3/31/2013 - Test Strip Lot # 19006M;    Kit Lot # P212420 - Kit Expiration Date 4/30/2013 - Test Strip Lot # 1X009M;    Kit Lot # P212421 - Kit Expiration Date 4/30/2013 - Test Strip Lot # 1X009M;    Kit Lot # P210194 - Kit Expiration Date 4/30/2013 - Test Strip Lot # 1X010M;   Kit Lot # P217043 - Kit Expiration Date 5/31/2013 - Test Strip Lot # 1Y012M;    Kit Lot # P234499 - Kit Expiration Date 11/30/2013 - Test Strip Lot # 25011M;  Kit Lot # P234664 - Kit Expiration Date 11/30/2013 - Test Strip Lot # 26012M;  Kit Lot # P238025 - Kit Expiration Date 12/31/2013 - Test Strip Lot # 26012M;  Kit Lot # P238028 - Kit Expiration Date 12/31/2013 - Test Strip Lot # 26013M;  Kit Lot # P243938 - Kit Expiration Date 12/31/2013 - Test Strip Lot # 26012M."
    }
  ]
}