  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70260",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "address_2": "N/A",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-1053-2015",
      "product_description": "SmartSite Low Sorbing Infusion Set, Model No. 10015862  The SmartSite Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a SmartSite bag access port, a non-vented drip chamber, roller clamp, infusion pump segment, a male luer, and low sorbing tubing.",
      "product_quantity": "7,440 units",
      "reason_for_recall": "CareFusion is recalling the SmartSite Low Sorbing Infusion set due to disconnection and leakage issues.",
      "recall_initiation_date": "20150113",
      "center_classification_date": "20150204",
      "termination_date": "20160315",
      "report_date": "20150211",
      "code_info": "Lot No. 12056016, 12057027, and 12107105."
    }
  ]
}