  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91491",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1047-2023",
      "product_description": "MEDLINE UNITE FOOT & ANKLE DRILL BIT, 1.3MM Solid Core, REF MPN10013",
      "product_quantity": "190 units",
      "reason_for_recall": "The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.",
      "recall_initiation_date": "20221222",
      "center_classification_date": "20230201",
      "report_date": "20230208",
      "code_info": "GTIN 10888277699113, Lot Number ?"
    }
  ]
}