      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91491",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1046-2023",
      "product_description": "MEDLINE UNITE FOOT & ANKLE Screw, Polyaxial Locking, Mini T8 (2.0x8mm), REF MMSL2008",
      "product_quantity": "420 units",
      "reason_for_recall": "The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.",
      "recall_initiation_date": "20221222",
      "center_classification_date": "20230201",
      "report_date": "20230208",
      "code_info": "GTIN 10193489120707, Lot Number 810304"
    }
  ]
}