  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96110",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1040-2025",
      "product_description": "Medline procedure kits labeled as:    1) Port Access Kit, Reorder number CVI5130;   2) VANTEX 7FR 3L 20CM CVC STOCK+ BUNDLE US. Reorder number STCVC01USGV; general surgical tray",
      "product_quantity": "749 units",
      "reason_for_recall": "Certain Medline kits were manufactured and distributed with Procedure Masks with SO SOFT Earloops, and SOFT TOUCH II Surgical Masks which were subsequently recalled for not meeting the bioburden testing requirements of EN 14683.",
      "recall_initiation_date": "20241118",
      "center_classification_date": "20250128",
      "report_date": "20250205",
      "code_info": "1) CVI5130:  UDI/DI 10653160372349 (each) 00653160372342 (case), Lot Numbers:  2024040490, 2024073190, 2024082790, 2024101790,;   2) STCVC01USGV:  UDI/DI 10653160376965 (each) 00653160376968 (case), Lot Numbers:  2024050990.",
      "more_code_info": ""
    }
  ]
}