  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Princeton",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98091",
      "recalling_firm": "Integra LifeSciences Corp. (NeuroSciences)",
      "address_1": "1100 Campus Rd",
      "address_2": "",
      "postal_code": "08540-6650",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).",
      "recall_number": "Z-1036-2026",
      "product_description": "Codman Microsensor Basic Kit. Catalog Number: 626631US.     Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.",
      "product_quantity": "1,301 units",
      "reason_for_recall": "Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.",
      "recall_initiation_date": "20251212",
      "center_classification_date": "20260108",
      "report_date": "20260114",
      "code_info": "Catalog Number: 626631US. UDI-DI: 10381780514473. Lot Numbers: 7288007, 7288040, 7294017, 7305988, 7305989, 7305990, 7305991, 7307683, 7325489, 7336088, 7336089, 7336090, 7305992, 7346223, 7391752, 7391751, 7406650, 7406651, 7406652, 7408825, 7408826, 7408827, 7456174, 7456175, 7456176, 7456179, 7478039, 7478040, 7480073, 7480072, 7482162, 7503704, 7507495, 7503608, 7510153, 7510155, 7510154, 7534856, 7535008, 7535009, 7411809.",
      "more_code_info": ""
    }
  ]
}