{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lincoln",
      "state": "NE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98122",
      "recalling_firm": "Securitas Healthcare LLC",
      "address_1": "4600 Vine St",
      "address_2": "",
      "postal_code": "68503-2823",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution -US Nationwide and the country of Canada.",
      "recall_number": "Z-1032-2026",
      "product_description": "Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3",
      "product_quantity": "1898 units",
      "reason_for_recall": "Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.",
      "recall_initiation_date": "20251208",
      "center_classification_date": "20260108",
      "report_date": "20260114",
      "code_info": "All units of the Affected Device Models sold between January 1st, 2023 and June 26th, 2024.  UDI-DI: 00840948102660; Serial Numbers:000CCC0E6CC9 000CCC0E6CC6 000CCC0E6CCC 000CCC0E6CCE 000CCC0E6CC5 000CCC0E6CD0 000CCC0E6D1F 000CCC0E6D1B 000CCC0E6D24 000CCC0E6D1C 000CCC0E6D22 000CCC0E6D01 000CCC0E6D04 000CCC0E6D19 000CCC0E6D1A 000CCC0E6D1E 000CCC0E6CC8 000CCC0E6D23 000CCC0E6CCD 000CCC0E6D1D 000CCC0E6CCB 000CCC0E6CCA 000CCC0E6CC7 000CCC0E6D20 000CCC0E6CCF 000CCC0E627E 000CCC0E6286 000CCC0E628C 000CCC0E6282 000CCC0E627C 000CCC0E6279 000CCC0E6289 000CCC0E627A 000CCC0E628E 000CCC0E627D 000CCC0E6288 000CCC0E6280 000CCC0E6284 000CCC0E6DDB 000CCC0E6D2E 000CCC0E6D28 000CCC0E6DE0 000CCC0E6D27 000CCC0E6D2C 000CCC0E6D2F 000CCC0E6DE3 000CCC0E6D2D 000CCC0E6D30 000CCC0E6DE1 000CCC0E6DD9 000CCC0E6DDF 000CCC0E6D25 000CCC0E6DDE 000CCC0E6DE4 000CCC0E6D2A 000CCC0E6D2B 000CCC0E6D29 000CCC0E6DDC 000CCC0E6DDA 000CCC0E6D26 000CCC0E6D36 000CCC0E6D32 000CCC0E6D06 000CCC0E6D33 000CCC0E6D3C 000CCC0E6D38 000CCC0E6D3B 000CCC0E6D07 000CCC0E6D05 000CCC0E6D0B 000CCC0E6D0A 000CCC0E6D0C 000CCC0E6D03 000CCC0E6D37 000CCC0E6D31 000CCC0E6D39 000CCC0E6D35 000CCC0E6D3A 000CCC0E6D34 000CCC0E6D08 000CCC0E6341 000CCC0E6D02 000CCC0E6D09 000CCC0E5B9F 000CCC0E5BD5 000CCC0E5BE0 000CCC0E6719 000CCC0E676D 000CCC0E676E 000CCC0E6DE2 000CCC0E6DDD 000CCC0E6CBC 000CCC0E6CC0 000CCC0E6CBF 000CCC0E6CBB 000CCC0E6CB9 000CCC0E6CBA 000CCC0E6D6A 000CCC0E6CC2 000CCC0E6CBE 000CCC0E6D61 000CCC0E6D67 000CCC0E6D68 000CCC0E6D62 000CCC0E6CC1 000CCC0E6D65 000CCC0E6D63 000CCC0E6CC3 000CCC0E6D66 000CCC0E6CBD 000CCC0E6CC4 000CCC0E6D64 000CCC0E6D69 000CCC0E6D6C 000CCC0E6D6B 000CCC0E6483 000CCC0E6482 000CCC0E647D 000CCC0E62A2 000CCC0E6484 000CCC0E633D 000CCC0E647A 000CCC0E647B 000CCC0E6340 000CCC0E6480 000CCC0E6481 000CCC0E647E 000CCC0E633C 000CCC0E62B6 000CCC0E647F 000CCC0E647C 000CCC0E6342 000CCC0E6343 000CCC0E633A 000CCC0E6479 000CCC0E6271 000CCC0E6278 000CCC0E6270 000CCC0E635C 000CCC0E6354 000CCC0E6353 000CCC0E6275 000CCC0E6355 000CCC0E6374 000CCC0E6357 000CCC0E6273 000CCC0E635B 000CCC0E6276 000CCC0E6352 000CCC0E626F 000CCC0E6358 000CCC0E6274 000CCC0E635A 000CCC0E626D 000CCC0E6356 000CCC0E6351 000CCC0E62A0 000CCC0E62A3 000CCC0E62A5 000CCC0E62BD 000CCC0E62C0 000CCC0E629D 000CCC0E62B5 000CCC0E62A1 000CCC0E629F 000CCC0E62B8 000CCC0E62BE 000CCC0E62BB 000CCC0E62A6 000CCC0E62BC 000CCC0E62BA 000CCC0E62A8 000CCC0E62BF 000CCC0E6277 000CCC0E6261 000CCC0E626E 000CCC0E6359 000CCC0E6265 000CCC0E6272 000CCC0E6531 000CCC0E652D 000CCC0E652E 000CCC0E6535 000CCC0E62EF 000CCC0E6537 000CCC0E6532 000CCC0E6530 000CCC0E6536 000CCC0E6538 000CCC0E652F 000CCC0E6533 000CCC0E6534 000CCC0E64E4 000CCC0E64DD 000CCC0E64DB 000CCC0E64DF 000CCC0E64DE 000CCC0E64E1 000CCC0E6329 000CCC0E632A 000CCC0E64D9 000CCC0E632C 000CCC0E6325 000CCC0E6326 000CCC0E64E2 000CCC0E6322 000CCC0E64E0 000CCC0E64E3 000CCC0E6327 000CCC0E6328 000CCC0E632B 000CCC0E64DC 000CCC0E64DA 000CCC0E6324 000CCC0E6323 000CCC0E6321 000CCC0E6469 000CCC0E62F6 000CCC0E6465 000CCC0E62F1 000CCC0E62F4 000CCC0E62F9 000CCC0E646A 000CCC0E62F5 000CCC0E62F3 000CCC0E646C 000CCC0E6468 000CCC0E6466 000CCC0E62F7 000CCC0E62FC 000CCC0E6464 000CCC0E6462 000CCC0E62FA 000CCC0E6467 000CCC0E62F2 000CCC0E6461 000CCC0E62FB 000CCC0E646B 000CCC0E62F8 000CCC0E6463 000CCC0E648D 000CCC0E648E 000CCC0E6486 000CCC0E64A2 000CCC0E6488 000CCC0E6487 000CCC0E64A8 000CCC0E648C 000CCC0E64A3 000CCC0E648A 000CCC0E649E 000CCC0E648F 000CCC0E649F 000CCC0E64A5 000CCC0E6490 000CCC0E6485 000CCC0E64A0 000CCC0E64A4 000CCC0E64A1 000CCC0E649D 000CCC0E64A6 000CCC0E6489 000CCC0E648B 000CCC0E64A7 000CCC0E626C 000CCC0E6268 000CCC0E626B 000CCC0E6371 000CCC0E6267 000CCC0E6373 000CCC0E636A 000CCC0E636B 000CCC0E6266 000CCC0E636F 000CCC0E6263 000CCC0E6370 000CCC0E636D 000CCC0E626A 000CCC0E6264 000CCC0E6369 000CCC0E6269 000CCC0E636E 000CCC0E636C 000CCC0E62F0 000CCC0E62E6 000CCC0E62EC 000CCC0E62C2 000CCC0E62E8 000CCC0E62ED 000CCC0E62C4 000CCC0E62E5 000CCC0E6258 000CCC0E64CD 000CCC0E64D0 000CCC0E64CE 000CCC0E64D8 000CCC0E64D5 000CCC0E64D2 000CCC0E64D7 000CCC0E6491 000CCC0E64D3 000CCC0E649C 000CCC0E64CF 000CCC0E64D4 000CCC0E64D1 000CCC0E64D6 000CCC0E625E 000CCC0E6260 000CCC0E625F 000CCC0E625B 000CCC0E62C7 000CCC0E62EE 000CCC0E62C6 000CCC0E62C1 000CCC0E62C3 000CCC0E62C5 000CCC0E62E9 000CCC0E62CB 000CCC0E6259 000CCC0E61D4 000CCC0E6256 000CCC0E61D8 000CCC0E625A 000CCC0E61D6 000CCC0E625D 000CCC0E61D2 000CCC0E6257 000CCC0E61D5 000CCC0E61D7 000CCC0E61DA 000CCC0E625C 000CCC0E61D9 000CCC0E61DB 000CCC0E61D3 000CCC0E61D1 000CCC0E6255 000CCC0E61C6 000CCC0E61DC 000CCC0E649A 000CCC0E6498 000CCC0E649B 000CCC0E6493 000CCC0E6499 000CCC0E6497 000CCC0E6494 000CCC0E6492 000CCC0E6496 000CCC0E6495 000CCC0E633E 000CCC0E633F 000CCC0E6339 000CCC0E6344 000CCC0E633B 000CCC0E61CB 000CCC0E61CF 000CCC0E6230 000CCC0E6227 000CCC0E61C8 000CCC0E6226 000CCC0E6228 000CCC0E61C9 000CCC0E61C7 000CCC0E61C5 000CCC0E6229 000CCC0E61CE 000CCC0E622F 000CCC0E61CD 000CCC0E6141 000CCC0E6146 000CCC0E6144 000CCC0E6209 000CCC0E614B 000CCC0E6206 000CCC0E620A 000CCC0E6147 000CCC0E614C 000CCC0E6208 000CCC0E6143 000CCC0E6142 000CCC0E61F8 000CCC0E6145 000CCC0E6201 000CCC0E6205 000CCC0E620C 000CCC0E614A 000CCC0E6203 000CCC0E6202 000CCC0E620B 000CCC0E6207 000CCC0E6148 000CCC0E6204 000CCC0E6149 000CCC0E61DE 000CCC0E61E7 000CCC0E621F 000CCC0E6222 000CCC0E61E4 000CCC0E61DD 000CCC0E6220 000CCC0E6219 000CCC0E61E0 000CCC0E61E1 000CCC0E621C 000CCC0E61E5 000CCC0E6223 000CCC0E621B 000CCC0E6224 000CCC0E61E6 000CCC0E6221 000CCC0E621D 000CCC0E61E8 000CCC0E61DF 000CCC0E61E2 000CCC0E621A 000CCC0E61E3 000CCC0E621E 000CCC0E61EC 000CCC0E6250 000CCC0E61EA 000CCC0E61F3 000CCC0E6253 000CCC0E61EB 000CCC0E624D 000CCC0E6251 000CCC0E624F 000CCC0E624E 000CCC0E61ED 000CCC0E6249 000CCC0E61E9 000CCC0E624A 000CCC0E61F1 000CCC0E624C 000CCC0E61EF 000CCC0E61EE 000CCC0E61F0 000CCC0E61F2 000CCC0E6254 000CCC0E6252 000CCC0E624B 000CCC0E61C3 000CCC0E6213 000CCC0E6218 000CCC0E620F 000CCC0E61BB 000CCC0E6210 000CCC0E620D 000CCC0E61C1 000CCC0E6211 000CCC0E6216 000CCC0E61BE 000CCC0E61B9 000CCC0E61C0 000CCC0E6217 000CCC0E6212 000CCC0E61BD 000CCC0E61BC 000CCC0E61BF 000CCC0E6214 000CCC0E620E 000CCC0E61C2 000CCC0E6244 000CCC0E623F 000CCC0E6240 000CCC0E6242 000CCC0E6233 000CCC0E6234 000CCC0E6239 000CCC0E623A 000CCC0E623E 000CCC0E6246 000CCC0E6243 000CCC0E6241 000CCC0E6238 000CCC0E623B 000CCC0E6247 000CCC0E623C 000CCC0E6235 000CCC0E6237 000CCC0E6248 000CCC0E623D 000CCC0E6245 000CCC0E6231 000CCC0E6236 000CCC0E6232 000CCC0E622C 000CCC0E61F4 000CCC0E6215 000CCC0E62E7 000CCC0E62EA 000CCC0E62CC 000CCC0E62C9 000CCC0E62CA 000CCC0E62EB 000CCC0E62C8 000CCC0E61FB 000CCC0E61B7 000CCC0E6021 000CCC0E6015 000CCC0E5DF9 000CCC0E5DF5 000CCC0E61FE 000CCC0E61B1 000CCC0E61B5 000CCC0E61FD 000CCC0E61AD 000CCC0E61B0 000CCC0E61AE 000CCC0E61B4 000CCC0E5E1E 000CCC0E5DFE 000CCC0E5DF7 000CCC0E5DF8 000CCC0E5E24 000CCC0E5E1B 000CCC0E62A4 000CCC0E62A7 000CCC0E62B9 000CCC0E629E 000CCC0E62B7 000CCC0E64BC 000CCC0E64BB 000CCC0E64B6 000CCC0E630C 000CCC0E6313 000CCC0E64B7 000CCC0E628F 000CCC0E6283 000CCC0E627B 000CCC0E6281 000CCC0E6287 000CCC0E628B 000CCC0E634A 000CCC0E62CD 000CCC0E6350 000CCC0E6349 000CCC0E634C 000CCC0E62CE 000CCC0E62D3 000CCC0E62CF 000CCC0E62D0 000CCC0E634D 000CCC0E62D1 000CCC0E62D7 000CCC0E62D6 000CCC0E6345 000CCC0E62D4 000CCC0E6348 000CCC0E62D8 000CCC0E6346 000CCC0E6347 000CCC0E634B 000CCC0E62D2 000CCC0E634E 000CCC0E62D5 000CCC0E628D 000CCC0E628A 000CCC0E627F 000CCC0E6285 000CCC0E6290 000CCC0E5F92 000CCC0E5F8D 000CCC0E5F8E 000CCC0E5F90 000CCC0E5F9B 000CCC0E5F96 000CCC0E5F93 000CCC0E5F95 000CCC0E5F91 000CCC0E5FA2 000CCC0E5F9C 000CCC0E5F9E 000CCC0E5F9D 000CCC0E5F9A 000CCC0E5FE6 000CCC0E5F94 000CCC0E5FE9 000CCC0E5F97 000CCC0E5F98 000CCC0E5FE3 000CCC0E5F8F 000CCC0E5FE7 000CCC0E5E23 000CCC0E5E21 000CCC0E5E00 000CCC0E5DFB 000CCC0E5E1D 000CCC0E5E1C 000CCC0E5DFC 000CCC0E5DFA 000CCC0E601E 000CCC0E600D 000CCC0E6018 000CCC0E5DFD 000CCC0E5E1F 000CCC0E5E22 000CCC0E5E1A 000CCC0E5DF6 000CCC0E5DFF 000CCC0E5E20 000CCC0E5E19 000CCC0E600B 000CCC0E601D 000CCC0E600C 000CCC0E6022 000CCC0E609A 000CCC0E5FE4 000CCC0E5FE8 000CCC0E5FEA 000CCC0E5FE1 000CCC0E5FEC 000CCC0E5FEB 000CCC0E5FE5 000CCC0E5FE2 000CCC0E601B 000CCC0E6020 000CCC0E6097 000CCC0E61F7 000CCC0E61B3 000CCC0E61B2 000CCC0E6200 000CCC0E61FC 000CCC0E61F6 000CCC0E61F9 000CCC0E61B8 000CCC0E61F5 000CCC0E61CC 000CCC0E622A 000CCC0E61D0 000CCC0E622D 000CCC0E622E 000CCC0E6225 000CCC0E61CA 000CCC0E61B6 000CCC0E61FF 000CCC0E61FA 000CCC0E61AF 000CCC0E61BA 000CCC0E61C4 000CCC0E5FC0 000CCC0E5FC2 000CCC0E5FC9 000CCC0E5FCD 000CCC0E5FBA 000CCC0E5FD0 000CCC0E5FBE 000CCC0E5FCF 000CCC0E5FBF 000CCC0E5FD2 000CCC0E5FD4 000CCC0E5FCB 000CCC0E5FD3 000CCC0E5FC1 000CCC0E5FCE 000CCC0E5FB9 000CCC0E5FCC 000CCC0E5FB7 000CCC0E5FBB 000CCC0E5FD1 000CCC0E5FBC 000CCC0E5FBD 000CCC0E5FB8 000CCC0E5FCA 000CCC0E5DD9 000CCC0E5E17 000CCC0E5DD2 000CCC0E5DDA 000CCC0E5E0D 000CCC0E5E12 000CCC0E5DDB 000CCC0E5DD7 000CCC0E5DD5 000CCC0E5E18 000CCC0E5DD3 000CCC0E5E13 000CCC0E5E10 000CCC0E5E0E 000CCC0E5DDC 000CCC0E5E16 000CCC0E5E15 000CCC0E5DD6 000CCC0E5E14 000CCC0E5E0F 000CCC0E5DD1 000CCC0E5E11 000CCC0E5DD4 000CCC0E5DD8 000CCC0E5FF3 000CCC0E5FF0 000CCC0E5FED 000CCC0E60B0 000CCC0E5FF4 000CCC0E5FEE 000CCC0E5FF2 000CCC0E5FF6 000CCC0E60AF 000CCC0E5FF1 000CCC0E60B2 000CCC0E60AD 000CCC0E60AA 000CCC0E60AE 000CCC0E60B1 000CCC0E60A9 000CCC0E60A7 000CCC0E5FF7 000CCC0E5FF5 000CCC0E60AC 000CCC0E60AB 000CCC0E60A8 000CCC0E5FF8 000CCC0E5FEF 000CCC0E609B 000CCC0E608B 000CCC0E6083 000CCC0E60A2 000CCC0E60A4 000CCC0E608C 000CCC0E609D 000CCC0E60A3 000CCC0E608D 000CCC0E60A5 000CCC0E6088 000CCC0E6087 000CCC0E60A6 000CCC0E609C 000CCC0E608E 000CCC0E6089 000CCC0E6086 000CCC0E609F 000CCC0E60A1 000CCC0E6084 000CCC0E609E 000CCC0E608A 000CCC0E60A0 000CCC0E6085 000CCC0E6014 000CCC0E6017 000CCC0E6019 000CCC0E601A 000CCC0E6016 000CCC0E6010 000CCC0E601C 000CCC0E601F 000CCC0E6013 000CCC0E600E 000CCC0E600F 000CCC0E6012 000CCC0E6011 000CCC0E5FA3 000CCC0E5FA4 000CCC0E6099 000CCC0E6095 000CCC0E6092 000CCC0E5FA0 000CCC0E5FA1 000CCC0E5F99 000CCC0E5F9F 000CCC0E608F 000CCC0E6098 000CCC0E6091 000CCC0E6093 000CCC0E5FC4 000CCC0E5FC7 000CCC0E6049 000CCC0E6052 000CCC0E5FB5 000CCC0E5FC6 000CCC0E5FB2 000CCC0E6051 000CCC0E604A 000CCC0E6048 000CCC0E5FC3 000CCC0E604C 000CCC0E5FB3 000CCC0E5FB6 000CCC0E604B 000CCC0E604D 000CCC0E604F 000CCC0E5FC5 000CCC0E6047 000CCC0E5FB1 000CCC0E604E 000CCC0E5FC8 000CCC0E5FB4 000CCC0E6050 000CCC0E5EFF 000CCC0E5F05 000CCC0E5EFE 000CCC0E6090 000CCC0E6094 000CCC0E6096 000CCC0E5DCF 000CCC0E5DE5 000CCC0E5DE2 000CCC0E5DC7 000CCC0E5DCC 000CCC0E5F0B 000CCC0E5F13 000CCC0E5F0C 000CCC0E5DC3 000CCC0E5DBA 000CCC0E5DC4 000CCC0E5F0D 000CCC0E5DBC 000CCC0E5F0E 000CCC0E5DBF 000CCC0E5DB9 000CCC0E5DC2 000CCC0E5F0F 000CCC0E5F12 000CCC0E5DC0 000CCC0E5F10 000CCC0E5DBD 000CCC0E5DC1 000CCC0E5F14 000CCC0E5F0A 000CCC0E5F09 000CCC0E5F11 000CCC0E5DBE 000CCC0E5DBB 000CCC0E5DEE 000CCC0E5F4D 000CCC0E5F50 000CCC0E5DEF 000CCC0E5DEA 000CCC0E5F46 000CCC0E5998 000CCC0E5DE7 000CCC0E5DE8 000CCC0E5DE4 000CCC0E5DC8 000CCC0E5DE6 000CCC0E5DD0 000CCC0E5DCA 000CCC0E5DDE 000CCC0E5DCD 000CCC0E5DE1 000CCC0E5DC6 000CCC0E5DCE 000CCC0E5DCB 000CCC0E5DE3 000CCC0E5DF2 000CCC0E5DEB 000CCC0E5DF0 000CCC0E5F4E 000CCC0E5F45 000CCC0E5DF1 000CCC0E5F4C 000CCC0E5F48 000CCC0E5F49 000CCC0E5DED 000CCC0E5F4B 000CCC0E5F47 000CCC0E5DF4 000CCC0E5DEC 000CCC0E5F4F 000CCC0E5DF3 000CCC0E5F4A 000CCC0E5DDD 000CCC0E5F04 000CCC0E5F61 000CCC0E5F00 000CCC0E5F07 000CCC0E5F60 000CCC0E5F5D 000CCC0E5F03 000CCC0E5F67 000CCC0E5F66 000CCC0E5F06 000CCC0E5F08 000CCC0E5F02 000CCC0E5E33 000CCC0E5E3B 000CCC0E5E35 000CCC0E5E37 000CCC0E5E34 000CCC0E5E38 000CCC0E5E3C 000CCC0E5E32 000CCC0E5E31 000CCC0E5E39 000CCC0E5E36 000CCC0E5E3A 000CCC0E5DC9 000CCC0E5DE0 000CCC0E5DDF 000CCC0E5DE9 000CCC0E5DC5 000CCC0E60EF 000CCC0E5EBA 000CCC0E60F5 000CCC0E5EB7 000CCC0E5EB9 000CCC0E5EB5 000CCC0E5FFE 000CCC0E60F3 000CCC0E60F1 000CCC0E5EBB 000CCC0E5EBE 000CCC0E60F9 000CCC0E60F6 000CCC0E5EB8 000CCC0E5EBF 000CCC0E60F7 000CCC0E60F0 000CCC0E5EBC 000CCC0E60F2 000CCC0E60F8 000CCC0E5EC0 000CCC0E60F4 000CCC0E5EBD 000CCC0E5FA9 000CCC0E5FAE 000CCC0E5FD6 000CCC0E5FA7 000CCC0E5FD8 000CCC0E5FAA 000CCC0E5FA8 000CCC0E5FDE 000CCC0E5FDF 000CCC0E5FAF 000CCC0E5FAB 000CCC0E5FDB 000CCC0E5FB0 000CCC0E5FAD 000CCC0E5FDD 000CCC0E5FA6 000CCC0E5FD5 000CCC0E5FA5 000CCC0E5FE0 000CCC0E5FDA 000CCC0E5FDC 000CCC0E5FAC 000CCC0E5FD9 000CCC0E5FD7 000CCC0E6078 000CCC0E6038 000CCC0E607B 000CCC0E6079 000CCC0E6032 000CCC0E6031 000CCC0E603A 000CCC0E6033 000CCC0E6030 000CCC0E6077 000CCC0E607E 000CCC0E6036 000CCC0E6080 000CCC0E6035 000CCC0E602F 000CCC0E6037 000CCC0E6039 000CCC0E607F 000CCC0E6081 000CCC0E6034 000CCC0E6082 000CCC0E607A 000CCC0E607C 000CCC0E607D 000CCC0E5E05 000CCC0E5E03 000CCC0E5E57 000CCC0E5E5E 000CCC0E5E01 000CCC0E5E55 000CCC0E5E07 000CCC0E5E09 000CCC0E5E5C 000CCC0E5E56 000CCC0E5E08 000CCC0E5E60 000CCC0E5E5F 000CCC0E5E0B 000CCC0E5E06 000CCC0E5E5D 000CCC0E5E0C 000CCC0E5E04 000CCC0E5E58 000CCC0E5E59 000CCC0E5E0A 000CCC0E5E5B 000CCC0E5E5A 000CCC0E5E02 000CCC0E5F15 000CCC0E5F5C 000CCC0E5F1A 000CCC0E5F20 000CCC0E5F58 000CCC0E5F1F 000CCC0E5F5B 000CCC0E5F17 000CCC0E5F59 000CCC0E5F54 000CCC0E5F51 000CCC0E5F1C 000CCC0E5F57 000CCC0E5F18 000CCC0E5F55 000CCC0E5F1D 000CCC0E5F53 000CCC0E5F56 000CCC0E5F5A 000CCC0E5F1E 000CCC0E5F19 000CCC0E5F16 000CCC0E5F52 000CCC0E5F1B 000CCC0E5E38 000CCC0E5E31 000CCC0E5E37 000CCC0E5E3A 000CCC0E5E36 000CCC0E5E35 000CCC0E5E3B 000CCC0E5E34 000CCC0E5E3C 000CCC0E5E32 000CCC0E5E39 000CCC0E5E33 000CCC0E5F68 000CCC0E5F65 000CCC0E5EFD 000CCC0E5F5F 000CCC0E5F62 000CCC0E5F64 000CCC0E5F63 000CCC0E5F5E 000CCC0E5F01 000CCC0E56AA 000CCC0E53CC 000CCC0E53D1 000CCC0E56AC 000CCC0E53CD 000CCC0E53D0 000CCC0E5377 000CCC0E56AB 000CCC0E53D2 000CCC0E53D4 000CCC0E53CE 000CCC0E536E 000CCC0E56A8 000CCC0E5371 000CCC0E5370 000CCC0E53CF 000CCC0E536F 000CCC0E53D6 000CCC0E53D5 000CCC0E56A9 000CCC0E5376 000CCC0E5997 000CCC0E596F 000CCC0E5973 000CCC0E5936 000CCC0E5994 000CCC0E5938 000CCC0E5972 000CCC0E5990 000CCC0E5974 000CCC0E5992 000CCC0E5932 000CCC0E598E 000CCC0E5970 000CCC0E5935 000CCC0E5971 000CCC0E592D 000CCC0E596D 000CCC0E53B1 000CCC0E53B3 000CCC0E596E 000CCC0E53AC 000CCC0E53B0 000CCC0E596A 000CCC0E5933 000CCC0E53A8 000CCC0E592F 000CCC0E53AD 000CCC0E53AE 000CCC0E53B2 000CCC0E596B 000CCC0E5930 000CCC0E53AA 000CCC0E5934 000CCC0E592E 000CCC0E53AB 000CCC0E596C 000CCC0E53AF 000CCC0E53A9 000CCC0E5969 000CCC0E598F 000CCC0E598D 000CCC0E5931 000CCC0E5996 000CCC0E5993 000CCC0E5991 000CCC0E5995 000CCC0E5937 000CCC0E5590 000CCC0E55AE 000CCC0E558D 000CCC0E5592 000CCC0E5681 000CCC0E567A 000CCC0E5679 000CCC0E55B0 000CCC0E567B 000CCC0E5678 000CCC0E55AC 000CCC0E55AF 000CCC0E56AD 000CCC0E56B1 000CCC0E55AD 000CCC0E55AB 000CCC0E5677 000CCC0E56B0 000CCC0E56AE 000CCC0E558F 000CCC0E558E 000CCC0E5591 000CCC0E53D7 000CCC0E56A7 000CCC0E53D3 000CCC0E56B2 000CCC0E56AF 000CCC0E5958 000CCC0E5959 000CCC0E5978 000CCC0E5955 000CCC0E5980 000CCC0E597F 000CCC0E45D6 000CCC0E45D7 000CCC0E45EA 000CCC0E45B4 000CCC0E45E3 000CCC0E45BB 000CCC0E45D3 000CCC0E45F4 000CCC0E45E5 000CCC0E45D5 000CCC0E45D4 000CCC0E45BF 000CCC0E45AF 000CCC0E45DA 000CCC0E45B5 000CCC0E45DF 000CCC0E45E1 000CCC0E45E2 000CCC0E45E6 000CCC0E45B6 000CCC0E45E0 000CCC0E45DD 000CCC0E45B7 000CCC0E45D8 000CCC0E557E 000CCC0E559A 000CCC0E557B 000CCC0E558B 000CCC0E559D 000CCC0E5585 000CCC0E559E 000CCC0E5583 000CCC0E559C 000CCC0E557C 000CCC0E5584 000CCC0E557D 000CCC0E5582 000CCC0E5589 000CCC0E5586 000CCC0E5580 000CCC0E5588 000CCC0E5581 000CCC0E558C 000CCC0E5599 000CCC0E5680 000CCC0E55B6 000CCC0E567C 000CCC0E557A 000CCC0E5595 000CCC0E55B1 000CCC0E5682 000CCC0E556F 000CCC0E55B4 000CCC0E5578 000CCC0E5596 000CCC0E5577 000CCC0E5593 000CCC0E5576 000CCC0E55B5 000CCC0E55B2 000CCC0E5579 000CCC0E5594 000CCC0E567F 000CCC0E5598 000CCC0E55B3 000CCC0E5597 000CCC0E567E 000CCC0E567D 000CCC0E594C 000CCC0E593B 000CCC0E5946 000CCC0E5949 000CCC0E5950 000CCC0E594E 000CCC0E59B7 000CCC0E593A 000CCC0E594B 000CCC0E593C 000CCC0E594F 000CCC0E59BB 000CCC0E59BA 000CCC0E59B8 000CCC0E594A 000CCC0E5939 000CCC0E59B9 000CCC0E59BC 000CCC0E5947 000CCC0E5948 000CCC0E5945 000CCC0E593E 000CCC0E594D 000CCC0E593D 000CCC0E595E 000CCC0E5968 000CCC0E5981 000CCC0E5987 000CCC0E5985 000CCC0E5983 000CCC0E5984 000CCC0E5988 000CCC0E598B 000CCC0E5986 000CCC0E5960 000CCC0E5967 000CCC0E595F 000CCC0E5964 000CCC0E5982 000CCC0E5989 000CCC0E5963 000CCC0E5961 000CCC0E5965 000CCC0E5962 000CCC0E598A 000CCC0E595D 000CCC0E598C 000CCC0E5966 000CCC0E5372 000CCC0E536C 000CCC0E5570 000CCC0E5373 000CCC0E5687 000CCC0E5683 000CCC0E5684 000CCC0E5574 000CCC0E568A 000CCC0E5689 000CCC0E5688 000CCC0E5575 000CCC0E5572 000CCC0E5573 000CCC0E5374 000CCC0E536D 000CCC0E5685 000CCC0E568B 000CCC0E5686 000CCC0E568D 000CCC0E568C 000CCC0E5571 000CCC0E5375 000CCC0E568E 000CCC0E5671 000CCC0E5698 000CCC0E5670 000CCC0E5695 000CCC0E5699 000CCC0E5697 000CCC0E5691 000CCC0E566D 000CCC0E5673 000CCC0E5692 000CCC0E566F 000CCC0E5675 000CCC0E5693 000CCC0E569A 000CCC0E568F 000CCC0E5676 000CCC0E566C 000CCC0E566E 000CCC0E5690 000CCC0E5694 000CCC0E5672 000CCC0E566B 000CCC0E5925 000CCC0E5929 000CCC0E5927 000CCC0E5922 000CCC0E5941 000CCC0E5921 000CCC0E5944 000CCC0E5928 000CCC0E593F 000CCC0E5923 000CCC0E59B1 000CCC0E5943 000CCC0E592A 000CCC0E59B3 000CCC0E5924 000CCC0E5942 000CCC0E59B6 000CCC0E592C 000CCC0E5926 000CCC0E5940 000CCC0E592B 000CCC0E59B4 000CCC0E59B2 000CCC0E59B5 000CCC0E595B 000CCC0E597E 000CCC0E5977 000CCC0E5952 000CCC0E595A 000CCC0E597C 000CCC0E595C 000CCC0E5979 000CCC0E5953 000CCC0E5954 000CCC0E597A 000CCC0E597B 000CCC0E5956 000CCC0E5976 000CCC0E5975 000CCC0E5957 000CCC0E5951 000CCC0E597D 000CCC0E559B 000CCC0E5674 000CCC0E5696 000CCC0E45E7 000CCC0E558A 000CCC0E557F 000CCC0E5587 000CCC0E4A27 000CCC0E4A23 000CCC0E4A29 000CCC0E499F 000CCC0E4A28 000CCC0E49A4 000CCC0E499E 000CCC0E4A25 000CCC0E4A2A 000CCC0E4836 000CCC0E49AC 000CCC0E49B0 000CCC0E483B 000CCC0E49AB 000CCC0E49A9 000CCC0E483F 000CCC0E4890 000CCC0E4840 000CCC0E483A 000CCC0E5563 000CCC0E556D 000CCC0E5568 000CCC0E53DC 000CCC0E5569 000CCC0E556C 000CCC0E5565 000CCC0E5567 000CCC0E53D8 000CCC0E53E3 000CCC0E556A 000CCC0E5564 000CCC0E53E0 000CCC0E53E2 000CCC0E53DD 000CCC0E556E 000CCC0E556B 000CCC0E53DB 000CCC0E53DE 000CCC0E5566 000CCC0E562B 000CCC0E55C0 000CCC0E55C2 000CCC0E5629 000CCC0E562D 000CCC0E562E 000CCC0E562A 000CCC0E55BE 000CCC0E55C1 000CCC0E4746 000CCC0E562C 000CCC0E55BD 000CCC0E55BF 000CCC0E467D 000CCC0E5545 000CCC0E5548 000CCC0E55E5 000CCC0E5547 000CCC0E5544 000CCC0E5540 000CCC0E55E1 000CCC0E55DB 000CCC0E55E4 000CCC0E5541 000CCC0E55DD 000CCC0E55E3 000CCC0E55E0 000CCC0E55DE 000CCC0E55E2 000CCC0E5543 000CCC0E554B 000CCC0E554A 000CCC0E55DF 000CCC0E55DC 000CCC0E5546 000CCC0E5549 000CCC0E5542 000CCC0E55A4 000CCC0E559F 000CCC0E55A0 000CCC0E55A1 000CCC0E55A3 000CCC0E55BB 000CCC0E55B9 000CCC0E55B8 000CCC0E55A2 000CCC0E55B7 000CCC0E55BA 000CCC0E55BC 000CCC0E53D9 000CCC0E53DF 000CCC0E53E1 000CCC0E53DA 000CCC0E55E6 000CCC0E55A9 000CCC0E5628 000CCC0E5559 000CCC0E555B 000CCC0E5560 000CCC0E5623 000CCC0E55A5 000CCC0E555D 000CCC0E5561 000CCC0E55A8 000CCC0E555A 000CCC0E5624 000CCC0E5557 000CCC0E5558 000CCC0E55A6 000CCC0E5627 000CCC0E555C 000CCC0E5562 000CCC0E555F 000CCC0E5625 000CCC0E555E 000CCC0E55AA 000CCC0E5626 000CCC0E55A7 000CCC0E49CC 000CCC0E49C1 000CCC0E4733 000CCC0E49CA 000CCC0E4780 000CCC0E5381 000CCC0E5378 000CCC0E554D 000CCC0E52D5 000CCC0E52D7 000CCC0E5551 000CCC0E5379 000CCC0E537A 000CCC0E537C 000CCC0E5554 000CCC0E5550 000CCC0E5552 000CCC0E52D4 000CCC0E5555 000CCC0E5553 000CCC0E554E 000CCC0E537F 000CCC0E5556 000CCC0E554C 000CCC0E537B 000CCC0E5380 000CCC0E537E 000CCC0E537D 000CCC0E55C5 000CCC0E55D8 000CCC0E55D1 000CCC0E55C4 000CCC0E55CE 000CCC0E55CB 000CCC0E55D9 000CCC0E55D6 000CCC0E55D2 000CCC0E55C9 000CCC0E55C3 000CCC0E55CC 000CCC0E55C7 000CCC0E55CD 000CCC0E55C6 000CCC0E55D5 000CCC0E55C8 000CCC0E55CA 000CCC0E55D0 000CCC0E55D4 000CCC0E55D7 000CCC0E55D3 000CCC0E55CF 000CCC0E55DA 000CCC0E554F 000CCC0E45D6 000CCC0E45FB 000CCC0E4671 000CCC0E46DF 000CCC0E45F9 000CCC0E466A 000CCC0E46DD 000CCC0E466D 000CCC0E4717 000CCC0E471A 000CCC0E4723 000CCC0E4673 000CCC0E46DC 000CCC0E46E0 000CCC0E46E1 000CCC0E4670 000CCC0E46E6 000CCC0E466B 000CCC0E46E3 000CCC0E46E4 000CCC0E4721 000CCC0E466C 000CCC0E4672 000CCC0E46DE 000CCC0E466E 000CCC0E473D 000CCC0E4779 000CCC0E4715 000CCC0E4669 000CCC0E46E5 000CCC0E46DB 000CCC0E466F 000CCC0E467A 000CCC0E475A 000CCC0E4679 000CCC0E467E 000CCC0E4677 000CCC0E475D 000CCC0E4678 000CCC0E4756 000CCC0E467B 000CCC0E475C 000CCC0E4676 000CCC0E4753 000CCC0E4675 000CCC0E45F8 000CCC0E467C 000CCC0E475B 000CCC0E4754 000CCC0E4759 000CCC0E474B 000CCC0E47DF 000CCC0E474C 000CCC0E4747 000CCC0E4749 000CCC0E47E1 000CCC0E47D7 000CCC0E47D8 000CCC0E47D9 000CCC0E4751 000CCC0E47DD 000CCC0E474F 000CCC0E474E 000CCC0E474D 000CCC0E47E0 000CCC0E4748 000CCC0E49FF 000CCC0E494C 000CCC0E4A04 000CCC0E4A01 000CCC0E49FE 000CCC0E494A 000CCC0E4A07 000CCC0E4A08 000CCC0E4A02 000CCC0E49FD 000CCC0E4A03 000CCC0E4954 000CCC0E4949 000CCC0E494F 000CCC0E4953 000CCC0E494E 000CCC0E4951 000CCC0E4A06 000CCC0E4952 000CCC0E4950 000CCC0E4A05 000CCC0E4A00 000CCC0E47A2 000CCC0E479F 000CCC0E4A11 000CCC0E479E 000CCC0E4891 000CCC0E48BD 000CCC0E48C4 000CCC0E48C2 000CCC0E49E0 000CCC0E49DF 000CCC0E494D 000CCC0E49BC 000CCC0E49BE 000CCC0E47B8 000CCC0E47B9 000CCC0E47BB 000CCC0E49BA 000CCC0E47BD 000CCC0E49B9 000CCC0E47BA 000CCC0E47B7 000CCC0E47B5 000CCC0E49B6 000CCC0E47B1 000CCC0E47AB 000CCC0E47A9 000CCC0E476F 000CCC0E4775 000CCC0E476B 000CCC0E47AA 000CCC0E47AF 000CCC0E476C 000CCC0E45E4 000CCC0E45B3 000CCC0E45E9 000CCC0E45B1 000CCC0E45EE 000CCC0E45DC 000CCC0E45C5 000CCC0E45B8 000CCC0E45BA 000CCC0E48BE 000CCC0E48BB 000CCC0E49DE 000CCC0E49DD 000CCC0E48BA 000CCC0E49E2 000CCC0E49DC 000CCC0E49E1 000CCC0E49D9 000CCC0E49DA 000CCC0E49E3 000CCC0E48C1 000CCC0E47AC 000CCC0E47B0 000CCC0E47AD 000CCC0E47A7 000CCC0E476D 000CCC0E4770 000CCC0E4773 000CCC0E4772 000CCC0E47AE 000CCC0E476A 000CCC0E476E 000CCC0E4741 000CCC0E45B2 000CCC0E45BE 000CCC0E4734 000CCC0E49C2 000CCC0E4737 000CCC0E4750 000CCC0E47DB 000CCC0E45F6 000CCC0E45B9 000CCC0E45D9 000CCC0E45F0 000CCC0E45F5 000CCC0E4A0B 000CCC0E479C 000CCC0E4A0D 000CCC0E47A5 000CCC0E47A3 000CCC0E4A13 000CCC0E47A4 000CCC0E4A10 000CCC0E4A0A 000CCC0E47A1 000CCC0E4A0E 000CCC0E479B 000CCC0E4787 000CCC0E45F2 000CCC0E45EC 000CCC0E45C6 000CCC0E45C1 000CCC0E45F3 000CCC0E45ED 000CCC0E45BC 000CCC0E45C3 000CCC0E47BC 000CCC0E49B7 000CCC0E49C0 000CCC0E49BB 000CCC0E49BD 000CCC0E4783 000CCC0E478D 000CCC0E478A 000CCC0E4784 000CCC0E49D5 000CCC0E4788 000CCC0E4782 000CCC0E4774 000CCC0E49D7 000CCC0E4789 000CCC0E49D8 000CCC0E4786 000CCC0E49D6 000CCC0E49CE 000CCC0E49CD 000CCC0E49D2 000CCC0E49D1 000CCC0E49D0 000CCC0E49D4 000CCC0E4785 000CCC0E475E 000CCC0E4792 000CCC0E4793 000CCC0E4790 000CCC0E4765 000CCC0E4794 000CCC0E4763 000CCC0E4760 000CCC0E4762 000CCC0E4795 000CCC0E4764 000CCC0E4791 000CCC0E4798 000CCC0E4767 000CCC0E4766 000CCC0E4768 000CCC0E475F 000CCC0E47A6 000CCC0E49BF 000CCC0E47B2 000CCC0E47B6 000CCC0E47B3 000CCC0E4796 000CCC0E4761 000CCC0E4797 000CCC0E4769 000CCC0E4A09 000CCC0E4A14 000CCC0E4A12 000CCC0E4A0C 000CCC0E479D 000CCC0E4A0F 000CCC0E47A0 000CCC0E4799 000CCC0E473B 000CCC0E477D 000CCC0E473F 000CCC0E4740 000CCC0E477A 000CCC0E473E 000CCC0E4776 000CCC0E4731 000CCC0E4735 000CCC0E49C7 000CCC0E49C4 000CCC0E49CB 000CCC0E49C3 000CCC0E472E 000CCC0E49C5 000CCC0E49C9 000CCC0E4778 000CCC0E473C 000CCC0E4781 000CCC0E4732 000CCC0E48C3 000CCC0E477E 000CCC0E4742 000CCC0E4739 000CCC0E49C8 000CCC0E49C6 000CCC0E477F 000CCC0E4745 000CCC0E4743 000CCC0E4839 000CCC0E4738 000CCC0E4730 000CCC0E473A 000CCC0E477B 000CCC0E4718 000CCC0E4722 000CCC0E4719 000CCC0E4724 000CCC0E472B 000CCC0E472D 000CCC0E4726 000CCC0E4716 000CCC0E4727 000CCC0E471C 000CCC0E471E 000CCC0E471D 000CCC0E472C 000CCC0E4720 000CCC0E4729 000CCC0E4728 000CCC0E471B 000CCC0E472A 000CCC0E471F 000CCC0E4725 000CCC0E4710 000CCC0E45FA 000CCC0E470C 000CCC0E45FC 000CCC0E4668 000CCC0E4660 000CCC0E4714 000CCC0E4711 000CCC0E4665 000CCC0E4664 000CCC0E4713 000CCC0E4667 000CCC0E4662 000CCC0E470F 000CCC0E4663 000CCC0E4666 000CCC0E470E 000CCC0E4661 000CCC0E470B 000CCC0E4712 000CCC0E470D 000CCC0E4837 000CCC0E483C 000CCC0E49AF 000CCC0E4841 000CCC0E483D 000CCC0E483E 000CCC0E49B2 000CCC0E49B3 000CCC0E49AE 000CCC0E49B1 000CCC0E4838 000CCC0E4A16 000CCC0E48BF 000CCC0E49AA 000CCC0E4A15 000CCC0E4A1F 000CCC0E4A19 000CCC0E49F9 000CCC0E49B4 000CCC0E49F2 000CCC0E4A18 000CCC0E4A20 000CCC0E49E4 000CCC0E48BC 000CCC0E49FC 000CCC0E4A2B 000CCC0E49F4 000CCC0E4A1C 000CCC0E49F3 000CCC0E49F5 000CCC0E4A17 000CCC0E49F8 000CCC0E49FA 000CCC0E4A1E 000CCC0E4A1B 000CCC0E49F1 000CCC0E49F6 000CCC0E4A1A 000CCC0E49AD 000CCC0E49DB 000CCC0E48C0 000CCC0E49A1 000CCC0E49A0 000CCC0E49A2 000CCC0E4A21 000CCC0E49A5 000CCC0E49A7 000CCC0E499D 000CCC0E4A2C 000CCC0E4A22 000CCC0E49A3 000CCC0E49A6 000CCC0E4A24 000CCC0E4A26 000CCC0E49A8 000CCC0E45C0 000CCC0E45BD 000CCC0E4744 000CCC0E4736 000CCC0E477C 000CCC0E472F 000CCC0E46E2 000CCC0E45B0 000CCC0E45DB 000CCC0E45C4 000CCC0E45EF 000CCC0E45EB 000CCC0E45F1 000CCC0E45C2",
      "more_code_info": ""
    }
  ]
}