  "results": [
    {
      "status": "Ongoing",
      "city": "Crumlin (North)",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93775",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of SC and OK.",
      "recall_number": "Z-0992-2024",
      "product_description": "RX Series Copper (Cu) Assay Ref. Number CU2340",
      "product_quantity": "3,053 kits (2 US, 3,051 OUS)",
      "reason_for_recall": "Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order.  If copper assay is run after total protein it could lead to falsely elevated copper results.",
      "recall_initiation_date": "20231129",
      "center_classification_date": "20240206",
      "report_date": "20240214",
      "code_info": "Catalogue Number/GTIN: CU2340 05055273201949; Lot/batch: 592278, 610661, 610669, 629021, 647201, 657944;",