  "results": [
    {
      "status": "Terminated",
      "city": "Somerville",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84639",
      "recalling_firm": "Ethicon, Inc.",
      "address_1": "Us Highway 22 West",
      "address_2": "N/A",
      "postal_code": "08876",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. US and Puerto Rico only.",
      "recall_number": "Z-0989-2020",
      "product_description": "Coated VICRYL (Polyglactin 910) Suture, 18\" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.",
      "product_quantity": "23688 units",
      "reason_for_recall": "Some sutures in the lot may exhibit a small section of braid fraying, which can lead to reduced tensile strength at that point.",
      "recall_initiation_date": "20200103",
      "center_classification_date": "20200204",
      "termination_date": "20210813",
      "report_date": "20200212",
      "code_info": "lot PGZ908"
    }
  ]
}