  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93722",
      "recalling_firm": "Covidien LP",
      "address_1": "15 Hampshire St",
      "address_2": "",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of CA, CO, FL, IL, TX.",
      "recall_number": "Z-0986-2024",
      "product_description": "AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm-  -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures  Model ABSTACK30",
      "product_quantity": "54 units",
      "reason_for_recall": "Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility",
      "recall_initiation_date": "20231116",
      "center_classification_date": "20240206",
      "report_date": "20240214",
      "code_info": "GTIN 20884523006527 Lot number: N3J1860Y",
      "more_code_info": ""
    }
  ]
}