  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Kortrijk",
      "state": "",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93712",
      "recalling_firm": "Barco N.V.",
      "address_1": "President Kennedypark 35",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of New Jersey.",
      "recall_number": "Z-0979-2024",
      "product_description": "Barco MDSC-8527 NXF, Part Number K9352421",
      "product_quantity": "2 units",
      "reason_for_recall": "Barco became aware of a problem with I2C cables in the manufacturing process of a limited batch of MDSC-85xx NXF products.  The functionality of the built-in MNA decoder is blocked.",
      "recall_initiation_date": "20231220",
      "center_classification_date": "20240202",
      "report_date": "20240214",
      "code_info": "UDI/DI 05415334043312",
      "more_code_info": ""
    }
  ]
}