{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98069",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.",
      "recall_number": "Z-0976-2026",
      "product_description": "Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609340 Software Version: N/A Product Description: Affixus 9 mm, Length 340 mm, Left, Trochanteric Nail Component: N/A",
      "product_quantity": "24",
      "reason_for_recall": "Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.",
      "recall_initiation_date": "20251202",
      "center_classification_date": "20251222",
      "report_date": "20251231",
      "code_info": "Lot Code: Model No 815609340 ; UDI-DI (01)00887868144486(17)340904(10)66717602 ; Lot Number 66717602 Model No 815609340 ; UDI-DI (01)00887868144486(17)340904(10)66892901 ; Lot Number 66892901"
    }
  ]
}