  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89860",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.",
      "recall_number": "Z-0972-2022",
      "product_description": "Artis zee biplane MN, Model Number 10094143",
      "product_quantity": "7 systems",
      "reason_for_recall": "If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message \"TUBE HOT, have a break\". Several minutes later the system will block X-ray to prevent further damage and display the message \"NO XRAY: TUBE TOO HOT\".",
      "recall_initiation_date": "20220223",
      "center_classification_date": "20220421",
      "report_date": "20220427",
      "code_info": "UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402",
      "more_code_info": ""
    }
  ]
}