      "status": "Terminated",
      "city": "Irving",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81929",
      "recalling_firm": "Abbott Laboratories, Inc",
      "address_1": "1921 Hurd Dr",
      "address_2": "N/A",
      "postal_code": "75038-4313",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and to countries of: CHINA, THAILAND, GUATEMALA, and EL SALVADOR.",
      "recall_number": "Z-0971-2019",
      "product_description": "ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31    The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.",
      "product_quantity": "24,995 units",
      "reason_for_recall": "Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation.",
      "recall_initiation_date": "20190102",
      "center_classification_date": "20190301",
      "termination_date": "20210630",
      "report_date": "20190313",
      "code_info": "(a) 7D70 21  Lot Number UDI Number  36338UN17 (01)00380740012922(17)190106(10)36338UN17  79865UN17 (01)00380740012922(17)190309(10)79865UN17  12921UN17 (01)00380740012922(17)190503(10)12921UN17  43618UN17 (01)00380740012922(17)190609(10)43618UN17  45337UN17 (01)00380740012922(17)190714(10)45337UN17  84994UN17 (01)00380740012922(17)191012(10)84994UN17  24856UN17 (01)00380740012922(17)191113(10)24856UN17  41652UN18 (01)00380740012922(17)200120(10)41652UN18  99639UN18 (01)00380740012922(17)200226(10)99639UN18  99661UN18 (01)00380740012922(17)200413(10)99661UN18  78493UN18 (01)00380740012922(17)200725(10)78493UN18  95220UN18 (01)00380740012922(17)200907(10)95220UN18  31428UN18 (01)00380740012922(17)201031(10)31428UN18  33784UN18 (01)00380740012922(17)200611(10)33784UN18      (b) 7D70-31   36339UN17 (01)00380740012939(17)190106(10)36339UN17  97389UN17 (01)00380740012939(17)190309(10)97389UN17  12922UN17 (01)00380740012939(17)190503(10)12922UN17  43616UN17 (01)00380740012939(17)190609(10)43616UN17  45338UN17 (01)00380740012939(17)190714(10)45338UN17  84993UN17 (01)00380740012939(17)190901(10)84993UN17  08788UN17 (01)00380740012939(17)191012(10)08788UN17  24857UN17 (01)00380740012939(17)191113(10)24857UN17  41651UN18 (01)00380740012939(17)200129(10)41651UN18  99640UN18 (01)00380740012939(17)200226(10)99640UN18  99662UN18 (01)00380740012939(17)200413(10)99662UN18  33783UN18 (01)00380740012939(17)200509(10)33783UN18  33785UN18 (01)00380740012939(17)200611(10)33785UN18  78859UN18 (01)00380740012939(17)200725(10)78859UN18  95341UN18 (01)00380740012939(17)200907(10)95341UN18  31429UN18 (01)00380740012939(17)201031(10)31429UN18"
    }
  ]
}