  "results": [
    {
      "status": "Ongoing",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98000",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "600 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-4835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.",
      "recall_number": "Z-0965-2026",
      "product_description": "LifeShield Drug Library Management (DLM);  LifeShield Infusion Safety Software Suite v2.2    List Number 17003-02",
      "product_quantity": "7 customers",
      "reason_for_recall": "For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.",
      "recall_initiation_date": "20251124",
      "center_classification_date": "20251219",
      "report_date": "20251231",
      "code_info": "version 2.2.1  UDI-DI 00887709127227   Issue affects customers who migrated from LifeShield DLM v2.1x to LifeShield DLM v2.2.1 and have any version of a Plum Duo v1.1 drug library that was approved or locked for review before the migration."
    }
  ]
}