  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93731",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0965-2024",
      "product_description": "MEDLINE BLOOD CULTURE COLLECTION KIT, Reorder Number DYNDH1714A",
      "product_quantity": "4320 units",
      "reason_for_recall": "Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Various Adult Blood and Blood Culture Kits. This recall is due to kits built using a Bactec component lot and expiration date that were not reflected on the kit insert. There is a risk of using an expired component if the user does not check the Bactec inside the kit for its expiration date.",
      "recall_initiation_date": "20231221",
      "center_classification_date": "20240201",
      "report_date": "20240207",
      "code_info": "UDI/DI 40195327347551 (case), 10195327347550 (unit), Lot Numbers:  23KBC138",
      "more_code_info": ""
    }
  ]
}