  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91339",
      "recalling_firm": "Getinge Usa Sales Inc",
      "address_1": "1 Geoffrey Way",
      "address_2": "N/A",
      "postal_code": "07470-2035",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide and Worldwide Distribution:  Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.",
      "recall_number": "Z-0960-2023",
      "product_description": "Getinge Flow-c Anesthesia System, Part No. 6887700",
      "product_quantity": "8.016 Total (7.766 OUS; 250 US)",
      "reason_for_recall": "Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.",
      "recall_initiation_date": "20221222",
      "center_classification_date": "20230126",
      "report_date": "20230201",
      "code_info": "Part #       UDI 6887700   07325710009765 Serial Numbers 4003-5078  Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.",
      "more_code_info": ""
    }
  ]
}