  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Hillsboro",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89771",
      "recalling_firm": "Acumed LLC",
      "address_1": "5885 NW Cornelius Pass Rd",
      "address_2": "",
      "postal_code": "97124-9370",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, \tCO, CT, FL, GA, HI, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, \tOH, OK, OR, PA, SC, SD, TN, TX, UT, WA, Puerto Rico and the country of South Africa.",
      "recall_number": "Z-0958-2022",
      "product_description": "4.0mm x 10mm Cannulated Screw, Short Thread; Part number 3005-40010",
      "product_quantity": "537 units",
      "reason_for_recall": "Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm  may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture, joint impingement or damage, or soft tissue damage.",
      "recall_initiation_date": "20220309",
      "center_classification_date": "20220420",
      "report_date": "20220427",
      "code_info": "All Batch number distributed between 22 AUG 2018 to 07 FEB 2022 . UDI : 10806378022869",
      "more_code_info": ""
    }
  ]
}