  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Parsippany",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64330",
      "recalling_firm": "Extremity Medical LLC",
      "address_1": "300 Interpace Pkwy Ste 410A",
      "address_2": "N/A",
      "postal_code": "07054-1148",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA including the states of Florida, Pennsylvania, New Hampshire, Oklahoma, California, Maryland, Arizona, Ohio, North Carolina, Texas, Virginia, Illinois, New York and Tennessee, and the countries of Australia, Belgium, Israel, Netherlands and United Kingdom.",
      "recall_number": "Z-0952-2013",
      "product_description": "Extremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-00006 - Non-Sterile.  Contained in Extremity Medical Surgical Kits including CompressX, Hallu.X, IO Fix and XMCP.  Used in the selection of screw size when used with appropriate depth gauge.",
      "product_quantity": "238 units",
      "reason_for_recall": "The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than intended.",
      "recall_initiation_date": "20130110",
      "center_classification_date": "20130315",
      "termination_date": "20130718",
      "report_date": "20130327",
      "code_info": "Catalog Number 101-00006; Lot Number 00921538."
    }
  ]
}