  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "GORT, Galway",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75846",
      "recalling_firm": "Natus Manufacturing Limited",
      "address_1": "IDA Business Park",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US including NJ, NC, TX, IL, VA, NY, FL, CA, MS, WA, IN, and MA. and Internationally to Canada and Costa Rica.",
      "recall_number": "Z-0951-2017",
      "product_description": "Natus Warmette Blanket Cabinet and Warmette Dual Cabinet  The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hospital linens; while, the Natus Warmette Dual Cabinet is designed to store and warm blankets and hospital linens in one compartment and fluids (irrigation and/or injection) in the other compartment within the storage times and warming temperatures recommended by the manufacturers of those items",
      "product_quantity": "31",
      "reason_for_recall": "The fuse blows during initial power up or after a period of time.  This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.",
      "recall_initiation_date": "20161015",
      "center_classification_date": "20170106",
      "termination_date": "20190624",
      "report_date": "20170118",
      "code_info": "Warmette Blanket Cabinet Serial Numbers:   10216709, 10216711, 10218559, 10219214, 10219337, 10219338, 10219339, 10219340, 10219387, 10219388, 10219869, 10219870, 10219871, 10219872, 10219873,  1028880.      Warmette Dual Cabinet Serial Numbers:   10216713, 10216714, 10218500, 10218501, 10218522, 10218526, 10218528, 10218982, 10219257, 10219389, 10219390, 10220534, 10220536, 10220537, 10220538.",
      "more_code_info": ""
    }
  ]
}