  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Littleton",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79272",
      "recalling_firm": "Medtronic Navigation, Inc.-Littleton",
      "address_1": "300 Foster St",
      "address_2": "N/A",
      "postal_code": "01460-2017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and  to the states of :  AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV., and to the countries of :  Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Luxembourg, Martinique, Mexico, Netherlands, New Zealand, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-0927-2018",
      "product_description": "O-arm¿ O2 interventional fluoroscopic x-ray system; Catalog Number: B1-700-02000    The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighting 60 lbs or greater and having an abdominal thickness of greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.",
      "product_quantity": "449",
      "reason_for_recall": "During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding metal during the disconnecting/reconnecting process which may result in an electrical short and a shock to the field service technician.",
      "recall_initiation_date": "20170930",
      "center_classification_date": "20180308",
      "termination_date": "20190502",
      "report_date": "20180314",
      "code_info": "Serial Numbers: 1065R  C0875  C0876  C0877  C0878  C0879  C0930  C0931  C0932  C0933  C0934  C0935  C0936  C0962  C0963  C0964  C0965  C0966  C0967  C0973  C0974  C0977  C0978  C0981  C0982  C0985  C0986R  C0989  C0990  C0993  C0994R  C0997  C0998  C1001  C1002  C1007  C1010  C1011  C1013  C1015  C1016  C1021  C1022  C1025  C1027  C1029  C1031  C1033  C1035  C1036  C1037  C1038  C1040  C1042  C1043  C1044  C1045  C1046  C1047  C1050  C1051  C1052  C1053  C1054  C1056  C1057  C1059  C1060  C1061  C1062  C1063  C1066  C1067  C1068  C1069  C1070  C1071  C1073  C1074  C1075  C1076  C1077  C1079  C1080  C1081  C1083  C1084  C1085  C1086  C1087  C1089  C1090  C1091  C1092  C1094  C1096  C1097  C1098  C1099  C1101  C1103  C1104  C1107  C1108  C1109  C1110  C1113  C1116  C1117  C1121R  C1123  C1124  C1125  C1127  C1128  C1129  C1131  C1132  C1133  C1134  C1135  C1139  C1140  C1141  C1142  C1143  C1144  C1145  C1146  C1147  C1148  C1149  C1151  C1152  C1153  C1154  C1156  C1157  C1158  C1160  C1161  C1162  C1163  C1165  C1166  C1167  C1168  C1169  C1170  C1171  C1172  C1173  C1174  C1175  C1176  C1177  C1178  C1179  C1180  C1181  C1182  C1185  C1186  C1187  C1188  C1189  C1191  C1192  C1193  C1194  C1195  C1197  C1198  C1199  C1200  C1201  C1203  C1204  C1205  C1206  C1208R  C1209  C1210  C1211  C1212  C1214  C1216  C1217  C1218  C1219  C1220  C1221  C1222  C1223  C1224  C1225  C1227  C1228  C1229  C1230  C1231  C1232  C1233  C1235  C1236  C1237  C1238  C1240  C1241  C1242  C1243R  C1244  C1245  C1246  C1247  C1249  C1251  C1252  C1253  C1254  C1255  C1256  C1258  C1260  C1261  C1262  C1263  C1264  C1265  C1266  C1267  C1268  C1269  C1270  C1271  C1274  C1275  C1276  C1277  C1278  C1279  C1280  C1281  C1282  C1283  C1284  C1285  C1286  C1288  C1289  C1290  C1291  C1292  C1293  C1294  C1295  C1296  C1297  C1298  C1299  C1300  C1301  C1302  C1303  C1304  C1305  C1306  C1307  C1308  C1309  C1311  C1313  C1314  C1315  C1316  C1317  C1318  C1319  C1320  C1321  C1322  C1323  C1324  C1325  C1326  C1327  C1328  C1329  C1330  C1331  C1332  C1334  C1335  C1336  C1337  C1338  C1339  C1340  C1341  C1342  C1343  C1344  C1345  C1346  C1347  C1348  C1349  C1350  C1351  C1352  C1353  C1354  C1355  C1356  C1357  C1358  C1359  C1360  C1361  C1362  C1363  C1364  C1365  C1366  C1367  C1368  C1369  C1370  C1371  C1372  C1373  C1374  C1375  C1376  C1377  C1378  C1379  C1380  C1381  C1382  C1383  C1384  C1385  C1386  C1387  C1388  C1389  C1390  C1391  C1392  C1393  C1394  C1395  C1396  C1397  C1398  C1399  C1400  C1401  C1402  C1403  C1404  C1405  C1406  C1407  C1408  C1409  C1410  C1411  C1412  C1413  C1414  C1415  C1416  C1417  C1418  C1419  C1420  C1421  C1422  C1423  C1424  C1425  C1426  C1427  C1428  C1429  C1430  C1431  C1432  C1433  C1434  C1435  C1436  C1437  C1438  C1439  C1440  C1441  C1442  C1443  C1444  C1446  C1447  C1448  C1449  C1450  C1452  C1453  C1454  C1455  C1457  C1458  C1459  C1460  C1461  C1462  C1463  C1464  C1465  C1467  C1468  C1469  C1470  C1471  C1472  C1473  C1474  C1475  C1476  C1477  C1478  C1479  C1481  C1482  C1483  C1484  C1485  C1486  C1488  C1489  C1490  C1491  C1492  C1493  C1494  C1495  C1497  C1498  C1499  C1500  C1501  C1502  C1503  C1504  C1505  C1506  C1507  C1508  C1509  C1510  C1511  C1512  C1513  C1514  C1515  C1516  C1517  C1518  C1519  C1520  C1521  C1522  C1523  C1524  C1525  C1526  C1527  C1528  C1529  C1530  C1531  C1532  C1533  C1534  C1535  C1536  C1537  C1538  C1539  C1540  C1541  C1542  C1543  C1544  C1545  C1546  C1547  C1548  C1549  C1550  C1551  C1552  C1553  C1554  C1555  C1556  C1557  C1558  C1559  C1560  C1561  C1562"
    }
  ]
}