{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Littleton",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64256",
      "recalling_firm": "Medtronic Navigation, Inc.",
      "address_1": "300 Foster St",
      "address_2": "N/A",
      "postal_code": "01460-2017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide and the countries of:  Canada, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CHILE, CHINA, CROATIA, CZECH REPUBLIC, DENMARK,FINLAND, FRANCE, GERMANY, GREECE, INDIA, IRELAND, ISRAEL, ITALY, JAPAN ,KOREA, REPUBLIC OF KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, MEXICO, NETHERLANDS, New Zealand, POLAND, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan(R.O.C.) , THAILAND, TURKEY, UNITED ARAB EMIRATES. and UNITED KINGDOM.",
      "recall_number": "Z-0926-2013",
      "product_description": "Medtronic O-arm Imaging System    Product Usage:  image-intensified fluoroscopic x-ray system, mobile",
      "product_quantity": "463 units",
      "reason_for_recall": "Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction",
      "recall_initiation_date": "20130131",
      "center_classification_date": "20130308",
      "termination_date": "20160706",
      "report_date": "20130320",
      "code_info": "Product Catalog Numbers:  9732719, 9732719R, 9733346, 9733346R, BI-700-00027, BI-700-00027R, BI-700-00027-GER, BI-700-00027-GERR, BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R    Units with serial numbers 101 through 564  and unit 580 are affected.  Serial numbers 565 and higher are not affected .",
      "more_code_info": ""
    }
  ]
}