  "results": [
    {
      "status": "Terminated",
      "city": "Strongsville",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84562",
      "recalling_firm": "Lumitex Inc",
      "address_1": "8300 Dow Cir",
      "address_2": "N/A",
      "postal_code": "44136-6600",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states: WI.",
      "recall_number": "Z-0923-2020",
      "product_description": "Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001",
      "product_quantity": "6837",
      "reason_for_recall": "There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.",
      "recall_initiation_date": "20191211",
      "center_classification_date": "20200131",
      "termination_date": "20210908",
      "report_date": "20200212",
      "code_info": "2014 Production Serial Numbers: All Serial numbers beginning with QAAT;   2015 Production Serial Numbers: All Serial numbers beginning with QAAU;   2016 Production Serial Numbers: All Serial numbers beginning with QAAV;   2017 Production Serial Numbers: All Serial numbers beginning with QAAW;   2018 Production Serial Numbers: Serial numbers QAAT60001 through QAAT60156"
    }
  ]
}