{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84497",
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "address_1": "800 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states:  MI and TX.",
      "recall_number": "Z-0920-2020",
      "product_description": "BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))",
      "product_quantity": "7",
      "reason_for_recall": "The gas spring on the front and rear protective cover may fail.  There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.",
      "recall_initiation_date": "20191209",
      "center_classification_date": "20200130",
      "termination_date": "20200917",
      "report_date": "20200205",
      "code_info": "Serial Numbers:  W1-6C000006,   W1-71000007,   W1-73000009,   W1-73000010,   XP-68000687T,   W1-71000008,   W1-85000011"
    }
  ]
}