{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75834",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: AZ, WA, CA, LA, FL, SC, and NJ.",
      "recall_number": "Z-0918-2017",
      "product_description": "Shimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems with 0.7/1.2JG326D X-Ray Tube    Diagnostic x-ray high voltage generator.",
      "product_quantity": "13 units",
      "reason_for_recall": "Shimadzu Medical Systems is recalling the Shimadzu X-ray High Voltage Generator due to the possibility of oil leaking from the x-ray tube housing under certain conditions.",
      "recall_initiation_date": "20161120",
      "center_classification_date": "20161223",
      "termination_date": "20171211",
      "report_date": "20170104",
      "code_info": "Serial No.  3M5249D1C013  3M5249D22013  3M5249D29002    3M5249D2A004  3M5249D2C004     3M5249D31009  3M5249D32012  3M5249D33003   3M5249D34001  3M5249D35002  3M5249D37006  3M5249D41006  3M5249D46003"
    }
  ]
}