  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75851",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT,NC, NE NH, NJ, NM, NY, NV, OH, OK, PA, RI, SC, SD, TN, TX,, VA, WA, WO, WY, WV.     OUS:     Algeria  Armenia  Australia  Austria  Brazil  Canada  Chile  China  Denmark  Egypt  France  Germany  Guatemala  Honduras  Hong Kong  Hungary  Iceland  India  Ireland  Italy  Japan  Korea  Libya  Mexico  Netherlands  Netherlands  Peru  Poland  Portugal  Russia  Saudi Arabia  Serbia  Sierra Leone  Slovakia  Spain  Taiwan  Thailand  Turkey  United Arab Emirates  United Kingdom  Venezuela  Vietnam",
      "recall_number": "Z-0913-2017",
      "product_description": "Prestige VH",
      "product_quantity": "Total for all products 1,598 (USA 824, OUS 774)",
      "reason_for_recall": "GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.",
      "recall_initiation_date": "20161121",
      "center_classification_date": "20161223",
      "termination_date": "20170926",
      "report_date": "20170104",
      "code_info": "Mfg. Lot or Serial # System ID  Not Available 708425PREST  00000063646YY4 314525APR  00000034238YY6 910249LRF3  00000023833YY7 919472CGRFA  00000063653YY0 724284R4  00000052120YY3 713442KPDRS  00000063577YY1 509458DR10  00000048069YY9 AH4317XR04  00000052117YY9 AH3221XR06  00000026253YY5 AH5681XR02  Not Available CS1029RX01  00000113920MO3 EG1112RX05  00000023862YY6 M4159446  00000034208YY9 M4160037  Not Available M1875113  00000039660YY6 M4500608  00000048049YY1 M2880713  00000026209YY7 M4046528  Not Available M2838405  Not Available M2663805  00000089618MO3 M4202921  00000031488YY0 M9624214  00000023814YY7 M4496929  00000023829YY5 M4496929  Not Available B4192927  00000034269YY1 HC2834XR11  00000063668YY8 HC1627XR14  00000026202YY2 HC0393XR02  00000023796YY6 HC0760XR01  00000026216YY2 PC0545XR04  Not Available 490314XR02  Not Available MC5880XR03  Not Available CC2149XR02  Not Available IS1002RX01  00000063686YY0 IE1012RX03  00000063667YY0 IE1025RX02  Not Available IE1006RX07  00000063609YY2 A5250603  Not Available A5127805  00000034283YY2 NO1004RX01  00000026203YY0 NO1041RX01  Not Available NO1104RX01  00000048004YY6 NO1016RX02  00000080076YY3 PPR00132  00000063713YY2 PPR59013  00000026248YY5 RU1306RX01  00000063560YY7 CE165306  00000063616YY7 05074FLU02  00000063755YY3 38853103XRMET"
    }
  ]
}