  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64000",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Argentina, Australia, Austria, Brazil, China, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Korea Republic, Malaysia, Netherlands, Russian Federation, Spain, Switzerland, Taiwan, Turkey, United Kingdom, and Viet Nam.",
      "recall_number": "Z-0913-2013",
      "product_description": "Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium.",
      "product_quantity": "28 in U.S.; 4 Canada; 318 to other foreign countries.",
      "reason_for_recall": "Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.",
      "recall_initiation_date": "20121128",
      "center_classification_date": "20130306",
      "termination_date": "20140224",
      "report_date": "20130313",
      "code_info": "All serial numbers and all software versions 4.05 and prior.    Part#    Item#  A94907 AU5811-02  A94912 AU5821-02  A94917 AU5831-02  A94922 AU5841-02  A94908 AU5811-03  A94913 AU5821-03  A94918 AU5831-03  A94923 AU5841-03  A94909 AU5811-04  A94914 AU5821-04  A94919 AU5831-04  A94924 AU5841-04  A94910 AU5811-06  A94915 AU5821-06  A94920 AU5831-06  A94925 AU5841-06"
    }
  ]
}