  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Carlsbad",
      "address_1": "2271 Cosmos Ct",
      "reason_for_recall": "Potential issues with ventilator navigation ring interface and liquid ingress:  1) The navigation ring may become inoperative or operate intermittently;  2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.",
      "address_2": "",
      "product_quantity": "3 systems",
      "code_info": "Model: V60  Part Number: R1053618  V60 Ventilator, Japan Opt: CFLEX,AVAPS    Serial Numbers:  100039155  100054624  100054738",
      "center_classification_date": "20210122",
      "distribution_pattern": "Worldwide distribution - US Nationwide including Guam and Puerto Rico;  Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.",
      "state": "CA",
      "product_description": "Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.",
      "report_date": "20210203",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Respironics California, LLC",
      "recall_number": "Z-0909-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "86697",
      "more_code_info": "",
      "recall_initiation_date": "20201102",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}