  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63671",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "address_2": "N/A",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Austria, Canada, Chile, Czech republic, Estonia,France, Germany, Hungary, Inida, Ireland, Italy, Japan, Libyan Arab Jamahari, Mexico, Netherlands, New Zealand, Norway , Oman, Panama, Peru, Poland, Qatar, Reunion, Russia, Saudi Arabia, South Africa, South Korea,Spain, Sweden, Swizerland, Thailand, UAE, and the UK.",
      "recall_number": "Z-0900-2013",
      "product_description": "Philips Digital Diagnost     Stationary radiographic system",
      "product_quantity": "360",
      "reason_for_recall": "When the operator for a wall stand view selects an \"image rotation\" different from default, or such image rotation is preset  in the examination database, then the resulting image may be rotated in the wrong direction.  When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.",
      "recall_initiation_date": "20121030",
      "center_classification_date": "20130301",
      "termination_date": "20130729",
      "report_date": "20130313",
      "code_info": "All Digital Diagnost systems with Eleva software version 3.0.x"
    }
  ]
}