  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86646",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.",
      "recall_number": "Z-0897-2021",
      "product_description": "Philips Respironics V60 Ventilator Part Number 1137292 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.",
      "product_quantity": "835 systems",
      "reason_for_recall": "Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.",
      "recall_initiation_date": "20201030",
      "center_classification_date": "20210122",
      "termination_date": "20240919",
      "report_date": "20210203",
      "code_info": "Model: V60 Ventilator, Part Number 1137276  V60 Ventilator, Intl Opt CFLEX, AVAPS 3.00 SW    Serial Numbers:  100286055  100286057  100286203  100286204  100286207  100289426  100289690  100289701  100289702  100289703  100289704  100289711  100291365  100291472  100291487  100291842  100309654  100309725  100309726  100309728  100311169  100311171  100314702  100315038  100318590  100318593  100318613  100319271  100319286  100319296  100319298  100319300  100329969  100329972  100329979  100337251  100339218  100340917  100341273  100345558  100345603  100369688  100377935  100284364  100284366  100286200  100286201  100286202  100286206  100288337  100289430  100289495  100289498  100289684  100289700  100291197  100291363  100291364  100291370  100291468  100291473  100291474  100291724  100291732  100291825  100291833  100291837  100309647  100309649  100309653  100317131  100317280  100320643  100321695  100321704  100329980  100329985  100329990  100341375  100376390  100378223  100378238  100291839  100341895  100343575  100343606  100343657  100343665  100343671  100343688  100345123  100345523  100345524  100345525  100345549  100345562  100345600  100345601  100345615  100354424  100354462  100355897  100356203  100356210  100356213  100356261  100356269  100356285  100356293  100356296  100356305  100356768  100358348  100366950  100367148  100367154  100367160  100367182  100289100  100289422  100289423  100289424  100289425  100289427  100289428  100289431  100289433  100289434  100289436  100289437  100289438  100289480  100289481  100289482  100289485  100289492  100289496  100289680  100289686  100289697  100289698  100289699  100289705  100289706  100289707  100289712  100289713  100291367  100291450  100291464  100291465  100291467  100291469  100291470  100291475  100291477  100291479  100291481  100291482  100291483  100291484  100291485  100291486  100291725  100291726  100291727  100291728  100291729  100291731  100291734  100291735  100291736  100291738  100291829  100291831  100291832  100291834  100291838  100291844  100309579  100309648  100314677  100314683  100314696  100314698  100314699  100314700  100314701  100314703  100314704  100314705  100314706  100314707  100314708  100314709  100314876  100315600  100315894  100316065  100316066  100316905  100316906  100316909  100316910  100316912  100316913  100316914  100317126  100317130  100317132  100317135  100317264  100317269  100317271  100317272  100317273  100317275  100317277  100317278  100317281  100317283  100317284  100317285  100317383  100318591  100318592  100319274  100319290  100319292  100319294  100319295  100319299  100319772  100319773  100319781  100320640  100320650  100321697  100321700  100321701  100321705  100321713  100329968  100329971  100329977  100329978  100329981  100329983  100329986  100329987  100329988  100329991  100329992  100330120  100330122  100330124  100330126  100330128  100330129  100337250  100337253  100337254  100337255  100337256  100338661  100338974  100338976  100338977  100338979  100339025  100339026  100339029  100339030  100339033  100339035  100339038  100339039  100339040  100339041  100339042  100339043  100339044  100339200  100339202  100339203  100339204  100339205  100339206  100339211  100339215  100339216  100339217  100339219  100339885  100340360  100340362  100340363  100340913  100340914  100340916  100340918  100341049  100341051  100341058  100341121  100341122  100341123  100341169  100341266  100341278  100341362  100341367  100341368  100341369  100341371  100341381  100341869  100341879  100341883  100341884  100341886  100341888  100341907  100341996  100342532  100342535  100342542  100342549  100342550  100342553  100342554  100342557  100342560  100342562  100342563  100342567  100342568  100343106  100343518  100343520  100343527  100343536  100343542  100343545  100343546  100343550  100343552  100343554  100343556  100343557  100343564  100343565  100343571  100343573  100343574  100343577  100343579  100343580  100343581  100343582  100343583  100343585  100343586  100343587  100343588  100343589  100343591  100343592  100343594  100343598  100343599  100343600  100343601  100343605  100343610  100343611  100343614  100343615  100343617  100343626  100343631  100343639  100343649  100343652  100343658  100343663  100343664  100343666  100343669  100343680  100343683  100343684  100343690  100343703  100343704  100343705  100343710  100343711  100344136  100344232  100344234  100344371  100345113  100345118  100345119  100345343  100345501  100345504  100345505  100345506  100345507  100345508  100345509  100345510  100345511  100345512  100345513  100345514  100345515  100345516  100345519  100345521  100345522  100345526  100345527  100345528  100345529  100345530  100345531  100345532  100345533  100345534  100345535  100345536  100345537  100345538  100345540  100345541  100345542  100345543  100345545  100345546  100345547  100345551  100345552  100345553  100345555  100345560  100345561  100345563  100345565  100345567  100345568  100345569  100345572  100345574  100345576  100345578  100345579  100345580  100345582  100345584  100345587  100345590  100345595  100345596  100345599  100345602  100345604  100345605  100345606  100345610  100345611  100345613  100345614  100345616  100345619  100345620  100345621  100345622  100345625  100345626  100345848  100346935  100346937  100346941  100346942  100346944  100346945  100346947  100351158  100353249  100354013  100354020  100354023  100354025  100354029  100354031  100354032  100354036  100354423  100354459  100354467  100354469  100354472  100354481  100358128  100358174  100358183  100358186  100358323  100366927  100366959  100367145  100367151  100367157  100367166  100367175  100367179  100367181  100367263  100367264  100367266  100371140  100371161  100371208  100371217  100371429  100371440  100371612  100375134  100375224  100375279  100375296  100375383  100375415  100375589  100375593  100375700  100375704  100375709  100375716  100375723  100375724  100375725  100375726  100375728  100375730  100375731  100375732  100375733  100375735  100375737  100375738  100375823  100375825  100375827  100375828  100375832  100375833  100375835  100375838  100375839  100375841  100375914  100375915  100375977  100375978  100375979  100375980  100375981  100375982  100375983  100375984  100375985  100375986  100375988  100375989  100375994  100375998  100376362  100376363  100376367  100376372  100376381  100376386  100376392  100376410  100377938  100377941  100378087  100378204  100378207  100378212  100378213  100378224  100378225  100378237  100378293  100378306  100378312  100378315  100378441  100378459  100378472  100378473  100378474  100378666  100378670  100378671  100378672  100378682  100378684  100378685  100378687  100378842  100378847  100378848  100378878  100378879  100378881  100378882  100378883  100378884  100378885  100378887  100378891  100378893  100378894  100378896  100378898  100378899  100378901  100378902  100378923  100378927  100378930  100378933  100378935  100378936  100378982  100337252  100341862  100345548  100345554  100345557  100345559  100345577  100309732  100325266  100341385  100343636  100343686  100366951  100366956  100367185  100367186  201040659  201040676  201040685  100369661  100369695  100316904  100316911  100317267  100317270  100345609  100350735  100350923  100351494  100351545  100351548  100351550  100354437  100354445  100356204  100356207  100356208  100356215  100358199  100370883  100370961  100370996  100371001  100371109  100371120  100371122  100371128  100371135  100371144  100371163  100371165  100371167  100371179  100371203  100371215  100371216  100371219  100371220  100371421  100371425  100371428  100371435  100371602  100371609  100371614  100371643  100371647  100371673  100371681  100371682  100371683  100371684  100371685  100371686  100376388  100376397  100377877  100377921  100377934  100378016  100378019  100378056  100378058  100378075  100378083  100378085  100378202  100378209  100378211  100378280  100378281  100378290  100378447  100378448  100378463  100378466  100378467  100378469  100378642  100378663  100378664  100378857  100378924  100387550  100388733  100389383  100389390  100389393  100389394  100389405  100389419  100389421  100389423  100389424  100389431  100389437  100389451  100389452  100389453  100389457  100389463  100389998  100390096  100390097  100390120  100390121  100390125  100390126  100390127  100390131  100390139  100390201  100390407  100390410  100390411  100390412  100390417  100390482  100390487  100390488  100390489  100390495  100390499  100390500  100390501  100390504  100391040  100391465  100391479  100391875  100391876  100391887  100391983  100392131  100392251  100392270  100319792  100320634  100320636  100322645  100329982  100338563  100338644  100338975  100339032  100339036  100339037  100339207  100339210  100339213  100339214  100339224  100343555  100343567  100343597  100343634  100343687  100344365  100345556  100346890  100346916  100346943  100355908  100355915  100356017  100356205  100356260  100356262  100356271  100356280  100356286  100356298  100369662  100378471",
      "more_code_info": ""
    }
  ]
}