{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63822",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..",
      "recall_number": "Z-0881-2013",
      "product_description": "Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260.  A component of the Computed Tomography X-Ray System.",
      "product_quantity": "39 units",
      "reason_for_recall": "Software anomaly.  Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project",
      "recall_initiation_date": "20121024",
      "center_classification_date": "20130226",
      "termination_date": "20150723",
      "report_date": "20130306",
      "code_info": "The EBW Serial Numbers subject to software updates are as follows:  8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127,  10331, 10348,  10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725,  11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957,  and 11979."
    }
  ]
}