  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97920",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of  Australia, Canada, India.",
      "recall_number": "Z-0878-2026",
      "product_description": "MAMMOMAT Inspiration;",
      "product_quantity": "6 units (3 US, 3 OUS)",
      "reason_for_recall": "There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.",
      "recall_initiation_date": "20251024",
      "center_classification_date": "20251203",
      "report_date": "20251210",
      "code_info": "Model Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 6734, 6694, 10121, 10309, 10336, 10342;",
      "more_code_info": ""
    }
  ]
}