  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Englewood",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97891",
      "recalling_firm": "Paragon 28, Inc.",
      "address_1": "14445 Grasslands Dr",
      "address_2": "",
      "postal_code": "80112-7062",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.",
      "recall_number": "Z-0871-2026",
      "product_description": "Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S",
      "product_quantity": "20",
      "reason_for_recall": "Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.",
      "recall_initiation_date": "20251013",
      "center_classification_date": "20251202",
      "report_date": "20251210",
      "code_info": "UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030",
      "more_code_info": ""
    }
  ]
}