  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89596",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "",
      "postal_code": "90502-1328",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV.    O.U.S.: N/A",
      "recall_number": "Z-0868-2022",
      "product_description": "MODEL: X-RAY TV SYSTEM SONIALVISION G4",
      "product_quantity": "176 systems",
      "reason_for_recall": "It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.",
      "recall_initiation_date": "20220110",
      "center_classification_date": "20220406",
      "report_date": "20220413",
      "code_info": "Model: SonialVision G4  Generator Number: D150-40  Serial Numbers:  41C35166C003  41C3536B4001  41C351683006  41C3516B3001  41C35169C002  41C351699002  41C351699001  41C3517AA001  41C35169B002  41C351671001  41C3516A3001  41C351661002  41C351671007  41C351683003  41C351671005  41C35179B004  41C351674001  41C35168B002  41C3517AA002  41C351685004  41C35165B002  41C3535A5001  41C353597001  41C35359C001  41C351674004  41C351685003  41C35164C001  41C351641001  41C351671004  41C351679001  41C3535A1001  41C3535A3001  41C351664004  41C351679006  41C351684004  41C351674002  41C3535A2002  41C351669002  41C351646002  41C35164A001  41C35359C002  41C351692001  41C35168B001  41C351653004  41C351664002  41C351673004  41C351663002  41C35164B001  41C351683001  41C351683002  41C351799001  41C351686002  41C351646001  41C351679005  41C351693003  41C351665002  41C35179A001  41C35165A003  41C35168A003  41C35179B005  41C35179B003  41C351671006  41C351689001  41C351799002  41C351641002  41C351671002  41C351676002  41C35165A001  41C351643001  41C35165A002  41C351655001  41C351695003  41C351647001  41C35166A002  41C351693002  41C351695001  41C351664005  41C351673003  41C35166C001  41C3535AC001  41C35166A001  41C35168B003  41C35169C001  41C3517B1001  41C3518B4002  41C351653002  41C3518B4001  41C35168A004  41C351665003  41C351678002  41C351684001  41C351683005  41C351693001  41C351671003  41C35179C001  41C351677001  41C35167A001  41C35165B001  41C351676001  41C351662001  41C35168C002  41C351668002  41C351679004  41C351677002  41C35179B001  41C351684005  41C351692003  41C351653003  41C351666002  41C351682002  41C351685001  41C351677003  41C353597002  41C3517A2002  41C351661001  41C35168C005  41C35168A002  41C35169A001  41C351663001  41C351692004  41C35168C004  41C351664001  41C351681001  41C351651002  41C351684006  41C3535A2001  41C351679007  41C351799003  41C351667001  41C351678001  41C351683004  41C351681002  41C351656001  41C351655003  41C3516B3002  41C3517B3002  41C3518B5001  41C3518B5002  41C3518B7001  41C3536B7001  41C351656002  41C35166C002  41C35168C001  41C351679002  41C351651001  41C351686001  41C351692002  41C3517A2001  41C35168A001  41C35166A003  41C351665001  41C351659001  41C3535A3002  41C351653005  41C351684003  41C3517A3001  41C351685002  41C3517A3002  41C351688001  41C351666001  41C351673002  41C351793001  41C351641003  41C351682001  41C351673001  41C35169B001  41C351695002  41C351655002  41C351679003  41C35168C003  41C35179B002  41C3517B1002  41C3517B3001  41C351668001  41C351669001  41C351654002",
      "more_code_info": ""
    }
  ]
}