{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.",
      "address_2": "",
      "product_quantity": "3",
      "code_info": "USN1800002  USN1800003  USN1800004",
      "center_classification_date": "20190221",
      "distribution_pattern": "Domestic distribution nationwide. Foreign distribution worldwide.",
      "state": "MA",
      "product_description": "PageWriter TC50 Government Bundle, Product 860429, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients",
      "report_date": "20190227",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips North  America, LLC",
      "recall_number": "Z-0860-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81957",
      "more_code_info": "",
      "recall_initiation_date": "20190109",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}