  "results": [
    {
      "status": "Ongoing",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81785",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "N/A",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide distribution to the following states AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the District of Columbia and Puerto Rico.",
      "recall_number": "Z-0821-2019",
      "product_description": "MOSAIQ¿ Oncology Information System and Sequencer    MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.",
      "product_quantity": "3,950 units total (1,747 US)",
      "reason_for_recall": "There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge.  Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates.",
      "recall_initiation_date": "20181213",
      "center_classification_date": "20190208",
      "report_date": "20190220",
      "code_info": "MOSAIQ Radiation Oncology version 1.0 and higher.  Multi-ACCESS version 6.1 and higher.      UDI  858164002084  858164002091  858164002107  858164002220  858164002237"
    }
  ]
}