  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Snoqualmie",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95910",
      "recalling_firm": "Spacelabs Healthcare, Inc.",
      "address_1": "35301 Se Center St",
      "address_2": "",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR.  OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.",
      "recall_number": "Z-0810-2025",
      "product_description": "Xhibit Telemetry Receiver with Software, Model 96280.  Intended to provide monitoring system with patient data.",
      "product_quantity": "434",
      "reason_for_recall": "Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.",
      "recall_initiation_date": "20241125",
      "center_classification_date": "20250107",
      "report_date": "20250115",
      "code_info": "Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246",
      "more_code_info": ""
    }
  ]
}