  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86985",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-0805-2021",
      "product_description": "Uroskop Omnia, Model no. 10094910 - Product Usage: a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).",
      "product_quantity": "6",
      "reason_for_recall": "After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time.",
      "recall_initiation_date": "20200909",
      "center_classification_date": "20210108",
      "termination_date": "20210416",
      "report_date": "20210120",
      "code_info": "Serial Nos. 3417  3414  3415  3418  3428  3416"
    }
  ]
}