  "results": [
    {
      "country": "United States",
      "city": "Center Valley",
      "address_1": "3773 Corporate Pkwy",
      "reason_for_recall": "The deflector may not move freely on the ventricular catheter.  The deflector comes pre-positioned at the 5 cm mark on the catheter.  In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle.  There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter",
      "address_2": "",
      "product_quantity": "11",
      "code_info": "Lots 4503489878, 4503792749, 4503682927, 4503508385, 4503789879, 4503800248 and 4503682935.",
      "center_classification_date": "20130212",
      "distribution_pattern": "Nationwide Distribution including MI, PA and TX.",
      "state": "PA",
      "product_description": "Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric Cathet, w/Stylet and RT-ANG Guide    Central Nervous System Shunt and Components intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.",
      "report_date": "20130220",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Aesculap, Inc.",
      "recall_number": "Z-0805-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63244",
      "termination_date": "20140108",
      "more_code_info": "",
      "recall_initiation_date": "20120604",
      "postal_code": "18034-8217",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}