  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63793",
      "recalling_firm": "The Anspach Effort, Inc.",
      "address_1": "4500 Riverside Drive",
      "address_2": "N/A",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the state of Massachusetts",
      "recall_number": "Z-0783-2013",
      "product_description": "Anspach Effort, Inc.,  98-0061 MA-15C  Bearing Sleeve with Removable Bur Guard.  A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.",
      "product_quantity": "8 devices",
      "reason_for_recall": "Contact  between the bur and bur guard could generate metal fragments that may or may not be visible to the  surgeon and can potentially remain in the surgical site.     The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.",
      "recall_initiation_date": "20120905",
      "center_classification_date": "20130204",
      "termination_date": "20140813",
      "report_date": "20130213",
      "code_info": "Lot number: 2000930",
      "more_code_info": ""
    }
  ]
}