{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Braintree",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79031",
      "recalling_firm": "Haemonetics Corporation",
      "address_1": "400 Wood Rd",
      "address_2": "N/A",
      "postal_code": "02184-2412",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, ME, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, and WI.",
      "recall_number": "Z-0778-2019",
      "product_description": "Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00    Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.",
      "product_quantity": "11,208 units",
      "reason_for_recall": "Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.",
      "recall_initiation_date": "20170802",
      "center_classification_date": "20190128",
      "termination_date": "20200806",
      "report_date": "20190206",
      "code_info": "Catalog number: 0291A-00, 0291E-00"
    }
  ]
}