  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95887",
      "recalling_firm": "Moximed, Inc.",
      "address_1": "46602 Landing Pkwy",
      "address_2": "",
      "postal_code": "94538-6420",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.",
      "recall_number": "Z-0762-2025",
      "product_description": "MISHA Knee System Implant Small, Left, REF: 2-1001.    MISHA Knee System Implant Small, Right, REF: 2-1002.",
      "product_quantity": "115",
      "reason_for_recall": "Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.",
      "recall_initiation_date": "20241202",
      "center_classification_date": "20241223",
      "report_date": "20250101",
      "code_info": "REF/UDI-DI/Lot(Expiration):  2-1001/00856047005795/23032901(29-Mar-25), 23092102(21-Sep-25), 24032601(26-Mar-26). 2-1002/00856047005801/23032902(29-Mar-25), 23092503(25-Sep-25), 24040801(8-Apr-26).",
      "more_code_info": ""
    }
  ]
}