  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93649",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.",
      "recall_number": "Z-0748-2024",
      "product_description": "SOLTIVE SuperPulsed Laser Fibers  Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS",
      "product_quantity": "82,200 total units (5 per box)",
      "reason_for_recall": "Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.",
      "recall_initiation_date": "20231201",
      "center_classification_date": "20240118",
      "report_date": "20240124",
      "code_info": "Models UDI/GTIN TFL-FBX150S 00821925043916 TFL-FBX200S 00821925043930 TFL-FBX365S 00821925043992 TFL-FBX550S 00821925044036 TFL-FBX940S 00821925044074 TFL-FBX150BS 00821925043879 TFL-FBX200BS 00821925043978  All lots affected"
    }
  ]
}