  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86886",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:   Canada,  Albania  Armenia  Australia  Austria  Belarus  Belgium  Bosnia And Herzegovina  Cyprus  Czech Republic  Egypt  Estonia  France  Germany  Greece  Hungary  Iceland  Israel  Italy  Japan  Jordan  Korea  Kuwait  Latvia  Libyan Arab Jamahiriya  Lithuania  Malaysia  Martinique  Morocco  Myanmar,   Netherlands  Norway  Pakistan  Poland  Russian Federation  Serbia  Singapore  Slovakia  Spain  Sweden  Switzerland  Taiwan  Turkey  United Arab Emirates  United Kingdom  Viet Nam",
      "recall_number": "Z-0747-2021",
      "product_description": "Covidien TA Auto Suture Vascular Stapler with DST Series Technology  30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue.  Item Number: TA30V3S",
      "product_quantity": "3255 units US; 3574 units OUS",
      "reason_for_recall": "May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis",
      "recall_initiation_date": "20201123",
      "center_classification_date": "20201229",
      "report_date": "20210106",
      "code_info": "Lot Numbers:  P0B1142MY, P0C1515MY, P0E0634MY, P0E1353MY, P0C0492MY,P0D1454MYP0E1352MY,P0G0601Y,  P0C0493MY    UDI: 10884521043145, 20884521043142"
    }
  ]
}