  "results": [
    {
      "status": "Ongoing",
      "city": "Temecula",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95707",
      "recalling_firm": "Abbott Vascular Inc",
      "address_1": "26531 Ynez Rd",
      "address_2": "",
      "postal_code": "92591-4630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC.  OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.",
      "recall_number": "Z-0731-2025",
      "product_description": "20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices).",
      "product_quantity": "3,673",
      "reason_for_recall": "Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.",
      "recall_initiation_date": "20241106",
      "center_classification_date": "20241220",
      "report_date": "20250101",
      "code_info": "UDI-DI: 08717648013614, Lots: 60538750, 60538751, 60539113, 60545329",
      "more_code_info": ""
    }
  ]
}