  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Altstatten Sg",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91174",
      "recalling_firm": "Icotec Ag",
      "address_1": "Industriestrasse 12",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of GA.",
      "recall_number": "Z-0728-2023",
      "product_description": "VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702",
      "product_quantity": "6 devices",
      "reason_for_recall": "During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.",
      "recall_initiation_date": "20221103",
      "center_classification_date": "20221219",
      "report_date": "20221228",
      "code_info": "UDI/DI:  (01) 07640164844589, Serial Numbers:  200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045",
      "more_code_info": ""
    }
  ]
}