  "results": [
    {
      "status": "Ongoing",
      "city": "Miami Lakes",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89559",
      "recalling_firm": "Heartware, Inc.",
      "address_1": "14400 Nw 60th Ave",
      "address_2": "N/A",
      "postal_code": "33014-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution throughout the United States.  OUS distribution in Argentina, Armenia, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea (South), Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritius, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.",
      "recall_number": "Z-0726-2022",
      "product_description": "HeartWare Ventricular Assist Device (HVAD) System",
      "product_quantity": "197,889 units",
      "reason_for_recall": "Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.",
      "recall_initiation_date": "20220127",
      "center_classification_date": "20220304",
      "report_date": "20220316",
      "code_info": "All lots/serials for the following product model numbers:  1440, 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR, 1407AU, 1407BR, 1407CA, 1407CA-CLIN, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US, 1440, 1650, 1650, 1650CA-CLIN, 1650DE"
    }
  ]
}