      "status": "Ongoing",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97610",
      "recalling_firm": "DJO, LLC",
      "address_1": "5919 Sea Otter Pl Ste 200",
      "address_2": "N/A",
      "postal_code": "92010-6750",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.",
      "recall_number": "Z-0722-2026",
      "product_description": "Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02",
      "product_quantity": "1,594",
      "reason_for_recall": "The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.",
      "recall_initiation_date": "20250925",
      "center_classification_date": "20251126",
      "report_date": "20251203",
      "code_info": "REF/UDI-DI/Lot: DP163CT02/00888912164382/190523, 171221. DP163CT03-BLK-L,XL/00888912164481/150221, 180821",