  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86817",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Ecuador, Egypt, Estonia, Finland,  France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam.",
      "recall_number": "Z-0720-2021",
      "product_description": "MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.",
      "product_quantity": "685 panel boxes (20 panels per box) = 13,700 panels",
      "reason_for_recall": "Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.",
      "recall_initiation_date": "20201202",
      "center_classification_date": "20201223",
      "termination_date": "20230103",
      "report_date": "20201230",
      "code_info": "Catalog Number: B1017-216  UDI Code: (01)15099590657550(17)210610(11)200610(10)2021-06-10   Lot Number: 2021-06-10"
    }
  ]
}