  "results": [
    {
      "status": "Terminated",
      "city": "Carrollton",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86803",
      "recalling_firm": "B Braun Medical, Inc.",
      "address_1": "1601 Wallace Dr Ste 150",
      "address_2": "N/A",
      "postal_code": "75006-6690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, PA, TN,TX, WA.",
      "recall_number": "Z-0717-2021",
      "product_description": "APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.",
      "product_quantity": "248 units",
      "reason_for_recall": "There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient.  This could result in a  a compounded drug that does not meet the specified accuracy range for an individual ingredient.",
      "recall_initiation_date": "20201123",
      "center_classification_date": "20201223",
      "termination_date": "20230314",
      "report_date": "20201230",
      "code_info": "All APEX Compounding System with Software Version 1.0.5.57 or earlier"
    }
  ]
}